Basic functions
16
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
5
Methods
The method selection window can be opened by pressing
. The available options are
shown in Figure 17. Press the appropriate icon to open the desired window.
Figure 17: Method selection window
Note: For measurements in the UV-range, please make sure to always cover the
cuvette slot. Otherwise result fluctuations cannot be excluded!
5.1
Basic functions
Pressing the
icon will open the basic functions menu. There are six different
measurement options to choose from.
5.1.1
Factor
The basic function “Factor” enables the determination of measurement results
through the multiplication of an absorbance with a user defined factor. After having
selected this method, a window will appear asking for the factor and the
wavelength for the measurement and calculation. Confirming with
opens the
measurement window. Inserting the zero solution and pressing
will initiate the zero
measurement. After the successful measurement of the zero solution, the sample solution
needs to be inserted. Pressing
will start the measurement process. After the
measurement has been completed, the measurement
result will be displayed in the result
window.
5.1.2
Standard
The basic function “Standard” enables the definition of a sample’s concentration on
basis of the concentration of a standard solution and a zero solution. After having
selected this method, a window will appear asking for the standard concentration
and the wavelength for the measurement. Confirming the entries by pressing
will open
the measurement window. Inserting the zero solution and pressing
will initiate the zero
measurement. Then, the standard
solution needs to be inserted. Pressing
will start the
sample measurement. After the standard solution the sample solution has to be inserted.
Pressing
will start the sample measurement. After the measurement has been