Procedures for colored or turbid samples
23
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
Test
Test tube for correction (value B)
0-01 Zirconium
Proceed as described in the instructions for Test
0-01, but
do not add NANOFIX R2
, seal, mix.
0-02, 0-03, 0-04, 0-05, 0-06,
0-08 Ammonium 3-2000
Procee
d as described in the instructions for test 0-02 / 0-03 /
0-04 / 0-05 / 0-06 / 0-08, but
do not add NANOFIX
R2
, close, mix.
0-07 AOX 3
Almost all colors and turbidities are destroyed under
test conditions and do not interfere. Resistant colors
and turbidities cause deviating results which cannot
be circumvented.
0-09 Lead 5
The original test already contains a correction.
0-14 Cadmium 2
Fill
empty
test tube with 4.0 mL sample, add
0.2 mL R2, close and mix.
0-15 Carbonate hardness 15
Open carbonate hardness test tube, add 4.0 mL
sample solution, close, mix and adjust to zero
(value B). Open test tube again, add
NANOFIX R2
,
close and shake well. Measure after 2 min (analytical
result = A
– B).
0-17 Chlorine/Ozone 2
0-18 Chlorine dioxide 5
Fill
empty
test tube with 4.0 mL sample for each
test.
0-19 Chloride 200
Open chloride test tube, add 1.0 mL sample solution
and 1.0 mL distilled water, close, mix.
0-21 Chloride 50
Open chloride test tube, add 4.0 mL sample solution
and 1.0 mL distilled water, close, mix.
0-24 Chromate 5
Fill
empty
test tube with 4.0 mL sample, add
0.2 mL R2, close, mix.
0-11, 0-12, 0-22, 0-23, 0-26,
0-27, 0-28, 0-29, 0-30, 0-33,
0-36, 0-38 COD 40
–60000
Almost all colors and turbidities are destroyed under
test conditions and do not interfere. COD resistant
colors and turbidities cause deviating results which
cannot be circumvented.
0-31 Cyanide 08
Proceed as described in the instructions for test 0-31,
but add 0.5 mL distilled water instead of 0.5 mL R3.
0-32 Anionic surfactants 4
Proceed as described in the instructions for test
0-32, but instead of 0.5 mL R2 add 0.5 mL distilled
water.
0-34 Kationic surfactants 4
Proceed as described in the instructions for test 0-34,
but instead of 0.5 mL R2 add 0.5 mL distilled water.
0-35 DEHA 1
Open DEHA test tube, add 4.0 mL sample, close
and mix.
0-37 Iron 3
No correction possible.
0-40 Fluoride 2
No correction possible.
0-41 Formaldehyde 8
Open formaldehyde test tube, add 2.0 mL sample
solution, close and mix.
0-43 Hardness 20
Open hardness test tube, add 0.2 mL sample
solution, close and mix.
0-44 Hardness Ca / Mg
No correction possible.
0-45 Potassium 50
Open potassium test tube, add 2.0 mL sample
solution, close and mix.
0-46 Formaldehyde 10
Fill
empty
test tube with 2.0 mL distilled water,
2.0 mL sample solution and 1.0 mL R2, close and
mix.
0-47 Nonionic surfactants 15
No correction possible.
0-49 Silver 3
Proceed as described in the instructions for test 0-49,
but add 0.5 mL DMSO instead of 0.5 mL R3.
0-50 Organic acids 3000
Proceed as described in the instructions for test 0-50,
but instead of 2.0 mL R5 add 2.0 mL distilled water.
0-52 Complexing agents 10
Fill
empty
test tube with 4.0 mL sample and 1.0 mL
distilled water, close and mix (analytical result =
A + B).
0-53 Copper 5
Proceed as described in the instructions for test 0-53,