Special methods
37
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
Figure 41: Re-measuring the sample of the standard series
The second method of correcting outliers would be to make a new preparation of the faulty
solution and re-measure it. Pressing
and selecting the option
“Repeat measurement” will
open a list box with the various samples (see Figure 41). By selecting the sample to be re-
measured and then inserting the sample after being prompted by the photometer, the value
in the measurement data will be replaced and the calibration function and statistical data will
be re-calculated.
To increase the accuracy of a method, where there is a linear relationship between
concentration and absorbance, each concentration of the series can be entered multiple
times and measured. It is important, however, not to measure each cuvette several times,
but rather, to make several individual experiments for each concentration.
Figure 42: Entering the properties of the calibrated special method
After finalization of the calibration function, the corresponding method can be created by
pressing
. A dialogue (see Figure 42) identical to the method design
dialogue will open so that you can enter the properties of the user-defined calibration (see