Color measurements
40
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
memory when the cuvette is removed or when the menu is exited
(Please note: This function
is only available in the "Open Mode")
.
Removing the cuvette or exiting the measurement menu by means of other icons in the task
bar will end the measuring process and the result will be stored in the measured value
memory of the device.
5.6
Color measurements
The spectrophotometer provides the possibility to determine different color types
and furthermore allows the comparison of these with stored color references.
After having selected this menu through
, one can decide via option buttons
to perform either a color measurement or a color comparison against a reference
(as far as a color reference is defined via the color analysis menu already, yet (figure 45). A
sample description
can be inserted in the field “title”. After selecting the button “color
measurement”, one can decide on the color type to be measured; the field “light source” and
“observer” are automatically filled in. These can always be manually changed. In this case,
however, the selection of the color number automatically indicates the preference CIE L*a*b.
Confirming with
opens the measurement window.
Figure 46: Color measurement menu
Note: Please make sure to always cover the cuvette slot during color measurement!
After measurement of the zero solution and the color sample, the result is displayed in the
result window.