Test number
53
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
PtCo/Hazen/
APHA
2°
C
10, 14, 20, 50
50
Iodine
2°
C
10, 14, 20, 50
10
Gardner
2°
C
10
10
Hess-Ives
-
-
10, 14, 20, 50
-
Yellowness index
2°
C
10, 14, 20, 50
-
Yellowness
index
E313
2°, 10°
D65, C
10, 14, 20, 50
-
Beer color EBC
2°
C
10
10
Beer color ASBC
2°
C
10
10
ICUMSA sugar color
-
-
10, 14, 20, 50
10, 20, 50
ADMI color
2°
D65
10, 14, 20, 50
50
ASTM color
2°
C
10, 14, 20, 50
32.5
Saybolt color
2°
C
10, 14, 20, 50
50
Ph. Eur.
2°
C
50
12, 16
Klett
2°
C
10, 50
-
Table 3: Allowable combinations of illuminants, observers and cuvette types
5.7
Test number
Besides opening the measurement menu of the cuvette tests via barcode
detection, it can also be opened by selecting the related test from a list of tube
tests or by entering the test number (Figure 58). Upon entering the three-digit test
number, the information for the last submethod to be selected will be shown (Figure 58). By
entering the three-digit test number plus the corresponding submethod number, the
information for that submethod can be viewed. Tap
to delete the entry. Confirming
with
will initiate the start of the selected method.
Figure 58: Opening a method entering a test number