Connectivity
68
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
Figure 81: Connectivity settings
6.5.2
RS232
In the RS232 settings (see Figure 82) the baud rate, bits and parity can be set. Tap on the
respective text field to open a list box with the available settings. Confirm with
to save
the settings.
Please note: In case of data transfer problems, make sure the connectivity
settings match the connected devices
(see Section 6.9.1).
Figure 82: RS232 settings
6.5.3
LAN
Press the
icon to open the overview of the LAN settings (Figure 83). If the device is part
of a DHCP network, an IP-address will automatically be assigned, which can be viewed here
along with the MAC address, subnet mask, gateway and the DNS-Servers. In case of a
network setup with static IP addresses, the selection button needs to be set
to “static”. The
respective fields will get editable and the necessary data can be entered.