Figure 92: Result of a standard measurement
Various types of sample information can be entered using the icons located below the result
window. Press the
icon to view other available options.
Via the “Properties” command,
new sample information can be added below the measurement menu (see Section
5.2.1
).
The “Do not save” command prevents the measurement result from being stored in the
instrument’s memory when the cuvette is removed or when the menu is exited
(Please note:
This function is only available in the "Open Mode")
.
Removing the cuvette or exiting the measurement menu by means of one of the other icons
in the task bar will end the measurement process and save the result in the IQC memory
(see Section 6.6.7) of the device.
Alternatively, a standard measurement can also be defined later as part of a measurement
process. Following a measurement using the standard menu of the spectrophotometer, press
the
icon in the upper right hand corner of the result window to define a measurement
as a standard measurement. A window will open for the selection of the desired IQC
standard (see Figure 93). After entering the LOT for test kit and standard via the text keypad
and confirming with
, the result will be rated as shown in Figure 92.
Figure 93: Measuring a standard using the measurement menu for cuvette tests