In order to enter a user-defined standard, the
icon must be pressed
in the “Select
standard” dynamic list (see Figure 94).
Figure 94: Selecting a standard
A text keypad will pop up for entering the name of the user-defined standard. After the first
measurement with the self-defined standard, the name entered here will appear in the
dynamic list for standard selection along with the test number and test name (see Figure 94
“User defined Standard” entry).
Figure 95: Creating a user-defined standard
Tap the entry
“Method,” and then select the method for which you want to define the
standard
from the “Select method” list that appears (see Figure 95). The submethod field can
be used to select submethods, provided that multiple submethods are available for a method.
Use the numeric keypad to enter the reference value and confidence interval for the standard
in the
“setpoint” and “margin” fields. Enter the LOT for test kit and standard via the text