IQC
77
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
keypad and confirm with
to open the measurement window. After inserting the
appropriate sample solution and confirming with
, the measurement will be carried out
and the result will be displayed as shown in Figure 92. The user-defined standard will not be
saved in the list of standards until the measurement has been carried out.
The results of the standard measurements can be viewed in the IQC memory (see Section
6.6.7) of the instrument.
6.6.3
Multiple determination
The analysis of a multiple determination increases the precision of a method and allows the
detection of outliers.
Press the
icon to open the window for selecting the desired method of the multiple
determination (see Figure 96). Tap the entry
“Method” to open the list of methods available
on the instrument. Now you can select the desired method from this list. The submethods
can be selected in the submethod field, provided that mutiple submethods are available for a
method. Use the alphanumeric keypad to enter the maximum permissible diffusion
percentage in the “Margin” field. This is calculated as part of the measurement on the basis
of the largest difference in readings between two measured values based on the average of
the measurements.
Figure 96: Entering the data for performing a multiple determination
Enter the number of measurements and the LOT of the test kit used via the text keypad and
then confirm with
to open the measurement window. The sample solutions must be
inserted in the order they are retrieved (see Figure 97). Confirm with
to start the
particular measurement process.