IQC
78
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
Figure 97: Measurement process of a multiple determination
After inserting and measuring all of the sample solutions, the result of the measurement will
be displayed as shown in Figure 98. The result will be given the rating
“Passed” (green) or
“Failed” (red) in the upper right hand corner. The measured value is compared against the
specified confidence interval. All of the measurement results are listed in sequence and
compared against both, the calculated and maximum diffusion.
Figure 98: Result of a multiple determination
In the example above, the average value of the measurements is 117 mg/L O
2
and the
greatest difference of two measurement results 1 mg/L O
2
. The resulting 0% spread,
therefore, is lower than the set limit of 10%; the quality objective is rated as having "Passed."
Press the
icon to view other available options.
Via the “Properties” command, new
sample information can be added below the measurement menu (see Section 5.2.1
). The
“Do not save” command prevents the measurement result from being stored in the
instrument’s memory when the cuvette is removed or when the menu is exited
(Please note:
This function is only available in the "Open Mode")
.