IQC
83
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
Figure 104: Overview of the measurements in the IQC memory
You have the option of printing the results, exporting them as CSV- or PNG file or deleting
them (
Please note: Deleting results in the user mode is only available when administrator
rights are given
”
)
. Furthermore, the IQC memory can be selected according to various
criteria. Figure 105 shows the selection according to the IQC type. The list box is opened by
tappin
g the entry “IQC type” in the IQC memory (see Figure 104). After selecting an IQC
type, the IQC memory is selected accordingly.
Figure 105: Selection of the IQC type
Various selection criteria can be applied simultaneously.
Pressing the
icon gives you the
option of deleting the selection or exporting the selected data as a CSV (Comma Separated
Value) file to an external mass storage device. A name for the export folder can be entered
on the text keypad. The file it contains can then be processed further in a spreadsheet
program. The file name has the following format:
SNDevice_IQC_results_YYYYMMDD_HH:MM:SS.csv