Tools
92
NANOCOLOR
®
UV
/
VIS
II and
NANOCOLOR
®
VIS
II V 1.1 / 10.2017
MACHEREY-NAGEL tube test boxes have a 2D barcode on the back, which contains the
LOT number of the test. When pressing the barcode icon in the LOT tracking menu, the 2D
barcode scanner will become active and the 2D barcode from the box can be scanned. Test
name, test numer, LOT number and expiration date are automatically entered into the list in
the LOT tracking menu.
In case a test without 2D barcode is used, the 2D barcode is damaged or user specific
method is used, the LOT information ca
n be entered manually by choosing the entry “Manual
entry” via the
icon. The respective test can be choosen from the list of all tests. The LOT
number and the expiration date need to entered manually. Confirming with ok saves the
entry in the list.
Figure 120: Manual entry of
List entries can be deleted when marking them followed by pressing
the
icon. Confirming
the popup with the safety questions, will delete the entry from the list. Expired LOTs can be
easily deleted ch
oosing the option “Remove expired LOTs via
the
icon.
When setting the checkbox “Remember LOT number” the device will remind the LOT
number, which has been assigned to a test kit after measurement. He next time the test kit is
measured, the latest entered LOT number will automatically be added to the sample
information. When removing the cuvette, the LOT information will be stored together with the
result in the result memory (see also section 7).
Setting the checkbox “Expiration warning” will trigger the popping up of an expiration
warning, when a LOT number is choosen, which has expired. The user can decide weather
to cancel the measurement or to proceed with the expired chemistry. When proceeding, the
information that an expired LOT has been used will be stored together with the result and
also will be exported to the CSV file.
The LOT number can be added to any measurement result after the measurement by
opening the
sample information “LOT”.
7
Memory
The measurement memory can store up to 5000 measurements and 100 scans or color
measurements.
Attention: When the maximum storage capacity has been reached, the oldest
measurement results will be overwritten. Before the memory becomes full, a
warning to save the results will appear in the home screen. Color references will
not be affected by the automatic overwriting process.
In addition to the measurement results, all other sample information that was entered will
also be saved. The sample information will be linked to the measurement results in a
tamperproof way (GLP conformity). The measurement results can be retrieved by pressing