49
Manual heating block
NANOCOLOR
®
VARIO C2
/
VARIO 4
EN, 08.2021
WARNING
Damaged cuvettes can break during the digestion process. Do not use cuvettes that are
damaged or were dropped. Make sure that the cuvettes are not damaged before inserting
them into the heating block.
Prepare the sample according to the instructions in the package insert and the pictogram
instructions (Figure 14).
H
G
F
E
D
C
B
A
1/1
Min
20–80 %
Max
100
380
1220
1500 mg/L O₂
www.mn-net.com
Methode / Method / Méthode / Método
595 / 605 / 620 nm
0291 100–1500 mg/L O₂
CSB 1500
COD / DCO / DQO
Test 0‑29
REF 985029
A037167
/
985029
/
09840
2.0 mL
20–25 °C
160 °C / 30 min
148 °C / 2 h
M
10 min
HKS 61
Figure 14: Example of pictogram instructions
13.2 Insert cuvettes
NOTICE
Contaminated cuvettes can damage the heating block. Clean the outside of the
cuvette before inserting it into the bores of the heating block.