10. Cleaning
Prior to each installation, top and bottom rubber feet need to cleaned with wet cloth soaked
with water .
1 . Take bottom rubber foot and wipe out the dirt attached to its bottom side of foot .
2 . Take top rubber foot and wipe out the dirt attached to its bottom side of foot . As shown in the picture
Note
: After cleaning, make sure rubber feet are cleaned and dry with no residue of dirt or dust left on
them . Also, ensure floor area is cleaned and free of obstacles .
Recommended Cleaning Instruction
The exterior of the Easy Fit System can be cleaned using a damp soapy cloth for general cleaning duties .
Please ensure the cloth is damp and not wet . Ensure the exterior of the device is dry after cleaning, dry using
a clean dry cloth .
Care should always be taken when cleaning around electrical components where these are
installed.
Disinfecting (if necessary)
Should the Easy Fit System require a more thorough clean, the use of the Actichlor disinfectant product,
which is widely available in tablet form and used throughout the healthcare industry, is recommended for
use to ensure a thorough clean .
• Please follow the manufacturers safety instruction for the use of the cleaning product before use to
ensure safe use for the operator and the patient .
• Ensure the cloth is
damp
before the cleaning process .
DO NOT
use a wet cloth over electrical
systems .
• Be careful not to let water ingress into the device as although the device is IP rated, it is not
water tight .
Application is through a clean soaked (but damp) cloth applied to wipe the device down
Used in the following dilutions to ensure an effective clean:
• Actichlor dissolvable chlorine tablets provide a concentration of 1000 ppm of available chlorine (0 .1%)
per 1 tablet
• 1 tablet (1 .7g formed tablet (x1)) will create a virucidal solution, diluted in 1 litre of water to provide
effective means to clean a “Dirty” device . This is also ideal for use after an outbreak of the Norovirus /
winter vomiting and can be used as a precaution against C .Diff . It is effective against viruses, bacteria,
spores, yeasts and moulds .
• The contact time against the outer components of the device should be for 5 minutes to prevent
any virucidal infections without a degradation to the functionality of the device . 5 minutes is a
recommended contact time . The device can withstand a longer contact period but the 5 minute
recommendation as a minimum must be followed to provide an effective cleaning regime .
• Blood spills should be dealt with by an increased concentration of the solution – please refer to the
instructions on the manufacturers product labelling .
Concentration limits for differing cleans will be shown on the manufacturers Achtochlor tablet container,
however, this is reflected in the table below: