Mackworth SA180 Pre-Sales Information 

 

For further information on the product and details summarised within this document please refer to the Mackworth SA180 User 
Manual. 
 

If you require larger print or videos 

 these are available on the Prism Website http://prismmedical.co.uk

 

 
 

 

Intended Use

 

 
For internal use only.

 

 

The Mackworth SA180 is a Standing Aid to be used by trained personnel. The Stand Aid is a medical device which 
carries out safe lifting and transfers of an individual from one resting surface to another (such as a shower chair to a 
bed). Prism Medical UK recommends that the transfer of a patient is fully risk assessed and conducted safely over a 

short distance only. Carrying out each transfer correctly for minimal strain or risk to the caregiver, while providing complete 
safety, dignity and comfort to the client. The Mackworth SA180 is suitable for patients in the SITTING position only, who have a 
degree of weight-bearing ability but require assistance to stand. By removing the foot tray, The Stand Aid can be used as a 
walking aid if competent and trained person completes an adequate risk assessment of the surroundings, and the patient is 
observed during use. The Mackworth SA180 is designed to support and promote safe patient handling and transfer for both the 
patient and carer. 
 
The Stand Aid is one of two components that makes this possible. The other component, the sling, is a specially designed fabric 
accessory that attaches to the Stand Aid by means of the boom Loop attachment and straps, and holds the patient during 
operation. 
 
Please refer to the user guides supplied with the sling and reference them while reviewing this manual.  
 
The functions of raising and lowering the boom, and opening/closing the legs are accomplished by pressing buttons on the hand 
control. 

 

 

Warning Notes 

Your Mackworth SA180 has been manufactured and tested to exceed BS EN 10535:2006. This does not mean that it 
can be used without care. ALL OPERATORS should have read the operating instructions and appreciate this warning 
section. 

 

1.

 

ALL Stand Aids are less stable on sloping surfaces. A 5-degree slope is the maximum permitted and then only with great 
care. 

2.

 

ALL Stand Aids are dangerous to the person being transferred when used recklessly or pushed at speed. 

 

Operating Environment 

The Mackworth SA180 is suitable for use within the professional health care facility environment as well as the home health care 
environment. The Stand Aid is not suitable for any special environments. 
 
The Stand Aid is not intended to be used in environments where there are rapid changes in the environmental temperature and 
humidity during intended use.

Summary of Contents for SA180

Page 1: ...Doc No 993065 Revision of document A Revision Date 25 05 21 SA180 Stand Aid Pre Sales Document...

Page 2: ...Slings 4 Detachable Parts 5 Attaching detaching the Battery 5 Technical Specification 6 Mackworth SA180 Dimensions 6 Specifications 7 Electrical Specifications 7 Environmental Storage and Operating C...

Page 3: ...gned to support and promote safe patient handling and transfer for both the patient and carer The Stand Aid is one of two components that makes this possible The other component the sling is a special...

Page 4: ...e been approved by Prism Medical UK for use in combination with the product We recommend the use of the Mackworth Healthcare or Care Ability manufactured sling range type B applied part to be utilised...

Page 5: ...ment and detachment of the Battery can be completed safely by following the following procedure Attach 1 Grab Hold of the Battery from the top position as shown in figure 5 and place your fingers onto...

Page 6: ...993065 Revision A P a g e 6 of 12 Technical Specification Mackworth SA180 Dimensions All dimensions below are shown in millimetres Figure 8 Figure 9...

Page 7: ...rical Specifications Battery type 2 x 12 volt Rechargeable sealed lead acid Battery capacity 24 V DC 1 x 2 9 Ah Charger rated input 230VAC 50 60Hz Charger rated output 27 4 29 0 VDC 0 8A Electric Shoc...

Page 8: ...ty is void To ensure the safety and continued good function of your Stand Aid routine service must be performed on your Mackworth SA180 Stand Aid Service should be completed by a Prism Medical UK appr...

Page 9: ...cleaning process Application is through a clean damp cloth applied to wipe the device down Use in the following dilutions to ensure an effective clean Actichlor dissolvable chlorine tablets provide a...

Page 10: ...993065 Revision A P a g e 10 of 12...

Page 11: ...993065 Revision A P a g e 11 of 12...

Page 12: ...UK for any errors or omissions Prism Medical operates a policy of continuous improvement Specifications and other data are subject to change without notice Manufacturer contact details Prism Medical U...

Reviews: