Specifications 

Dimensions 

 

Turning Circle 

1230mm 

Overall Height 

1720mm 

Overall Width 

680mm 

Ground Clearance 

33mm 

Front and Rear Castors Size 

100mm 

 

Operational Forces

 

Moving Force (with 180 kg) 

104.5 N 

Moving Force (with 0 kg) 

20 N 

Operation of the Brake  

35 N 

Handset  

4.5 N

 

E-Stop Button  

12 N 

Manual Leg Lever (with 0kg) 

80 N 

Table 5 

 

Electrical Specifications 

Battery type:

                           2 x 12-volt Rechargeable sealed lead acid. 

Battery capacity:

                    24 V DC, 1 x 2.9 Ah 

Charger rated input:

             230VAC 50/60Hz 

Charger rated output:

           27.4/29.0 VDC @ 0.8A 

 

Electric Shock Protection 

Charger:                                

Class II *Lift ‐ internal power source

 

Degree of shock protection:

 

Type B *Lift ‐ internal power source

 

Duty cycle:

                            10% Use 

 90% Rest (90 seconds use - 810 seconds rest) 

IP rating, control box:

         IPX4 

IP rating, battery: 

 IPX5 

IP rating, handset:              

 IPX4  

 

 
 
 
 
 
 
 

 
 
 
 
 
 
 

Weights: 

Safe Working Load (SWL) 

180 kg 

Full Product  

47 kg 

Base Assembly 

23 kg 

Mast & Boom Assembly  

24 kg 

Table 3

 

Table 4

 

Summary of Contents for SA180

Page 1: ...Doc No 993065 Revision of document A Revision Date 25 05 21 SA180 Stand Aid Pre Sales Document...

Page 2: ...Slings 4 Detachable Parts 5 Attaching detaching the Battery 5 Technical Specification 6 Mackworth SA180 Dimensions 6 Specifications 7 Electrical Specifications 7 Environmental Storage and Operating C...

Page 3: ...gned to support and promote safe patient handling and transfer for both the patient and carer The Stand Aid is one of two components that makes this possible The other component the sling is a special...

Page 4: ...e been approved by Prism Medical UK for use in combination with the product We recommend the use of the Mackworth Healthcare or Care Ability manufactured sling range type B applied part to be utilised...

Page 5: ...ment and detachment of the Battery can be completed safely by following the following procedure Attach 1 Grab Hold of the Battery from the top position as shown in figure 5 and place your fingers onto...

Page 6: ...993065 Revision A P a g e 6 of 12 Technical Specification Mackworth SA180 Dimensions All dimensions below are shown in millimetres Figure 8 Figure 9...

Page 7: ...rical Specifications Battery type 2 x 12 volt Rechargeable sealed lead acid Battery capacity 24 V DC 1 x 2 9 Ah Charger rated input 230VAC 50 60Hz Charger rated output 27 4 29 0 VDC 0 8A Electric Shoc...

Page 8: ...ty is void To ensure the safety and continued good function of your Stand Aid routine service must be performed on your Mackworth SA180 Stand Aid Service should be completed by a Prism Medical UK appr...

Page 9: ...cleaning process Application is through a clean damp cloth applied to wipe the device down Use in the following dilutions to ensure an effective clean Actichlor dissolvable chlorine tablets provide a...

Page 10: ...993065 Revision A P a g e 10 of 12...

Page 11: ...993065 Revision A P a g e 11 of 12...

Page 12: ...UK for any errors or omissions Prism Medical operates a policy of continuous improvement Specifications and other data are subject to change without notice Manufacturer contact details Prism Medical U...

Reviews: