background image

 

 

 

999562 – Revision A 

 

P a g e

 

5 of 28

 

1.5

 

Contraindications / Limitations 

There are no known “contraindications” associated with the usage of the TA400 Transfer Aid, provided they are used as per 

manufacturer’s recommendations and guidelines. However, it is recommended that a client specific assessment is completed by 

a trained and knowledgeable health care professional to determine the method of transfer. 

 

Mackworth does not recommend a required number of care givers for the use of our products. This information and 

recommendation can only be provided after a thorough personalised, case specific assessment, as there are many factors that 

can influence these decisions.

 

1.6

 

Intended Use 

For internal use only. 

 

The Mackworth TA400 is a sit to stand platform, which is designed to provide an active, safe and comfortable transfer of the 

user. It is ideal for short distance transfers, such as to/from bed to chair or from bedroom to bathroom. 

 

Its versatility allows use with wheelchairs, toilet and shower chairs. Soft and wide padding in the leg support improves stability, 

the sturdy handle offers grip options for both the user and the carer and also the padded seat paddles which can support the 

patient in a comfortable seated position. 

 

The rear wheels with independent locking manual brakes, makes it easy for a carer to manoeuvre, and provides maximum 

stability and safety for the user. 

 

The Mackworth TA400 has a safe working load of 180 kg (400 lb)

this load must not be exceeded. 

The device is used under instruction, and the operation of the aid is undertaken by a trained carer. 

A risk assessment must be performed before using any other manufactured Transfer Aid, to ensure safe use can be 

established. 

 

The safety belt is a specially designed fabric accessory that attaches to the Transfer Aid by means of buckle incorporated in the 

safety belt which clips onto the attachment pins either side of the handle, this holds and aids an individual while the transfer 

takes place. The safety belt is supplied separately from the Transfer Aid at the initial time of purchase.  

If additional accessories have been supplied with the Transfer aid, refer to the instructions included with those items. 

 

The Transfer Aid: 

 

Helps to assist a user with sit to stand transfers 

 

Helps to transfer penitent from one destination to another (i.e. bedroom to bathroom) 

 

Maximum user weight: 180 kg (400 lb) 

 

Padded leg support for added comfort 

 

Padded seat paddles for patient to sit while transferring form one destination to another 

 

Multiple grip options for both user and carer 

 

Safety Belt available to enable a carer to support the patient while a transfer is carried out 

 

Independent locking rear brakes 

 

 

The TA400 is intended to be installed on a flat and levelled surface prior to use.  

 

 

The Transfer Aid must be installed only by persons authorised by Mackworth or thoaw who have the rights to install 

and commission the Transfer Aid safe for use. 

 

 

Under no circumstance should the TA400 be put in control of a person who has not been properly trained in the use 

and care of this equipment. Failure to adhere to this warning may result in serious injury to the operator and/or the 

individual being transferred. 

 

 

The Transfer Aid, and associated belts are not toys. Do not use it for unsafe practices. Do not allow children to play with 

the Aid or any of its components. The TA400 should not be used for any practice except its intended use. 

Summary of Contents for TA400

Page 1: ...Doc No 999562 Revision of document A Revision Date 24 08 21 TA400 Transfer Aid User Manual...

Page 2: ...sitioning 9 4 2 To Alter the Leg Width 10 4 3 Operating Seat Pads 11 4 4 Applying the Brakes 11 4 5 Slings 12 4 5 1 Compatible Belt Types 12 Technical Specification 13 5 1 TA180 Dimensions 13 5 2 Spec...

Page 3: ...port it to the manufacturer and to your national authority Store this manual with the documents included with the Transfer Aid Contents of this manual are subject to change without prior written notic...

Page 4: ...ons before use Caution see instructions for use Manufacturer Safe Working Load represents the maximum load rated for safe operation Packaging indicator This way up Date of manufacture Serial number Pa...

Page 5: ...erformed before using any other manufactured Transfer Aid to ensure safe use can be established The safety belt is a specially designed fabric accessory that attaches to the Transfer Aid by means of b...

Page 6: ...way Protecting the people present visually monitor the clipped connection points during transfer stages so the belt remains firmly attached to the connection pins To reduce the risk of unintended use...

Page 7: ...s of the Transfer aid If you have not received all the components contact your Mackworth dealer immediately contact details are provided on the last page of this manual Item Description 1 Handle 2 Saf...

Page 8: ...a of the Transfer Aid The environment in which the Transfer Aid is situated in whilst carrying out a transfer is required to be on a flat surface with no steep inclines 3 2 Unboxing Transfer Aid When...

Page 9: ...ening or closing the seat pads Appropriate training in lifting and handling procedures should be undergone by any person operating a Transfer Aid for their own and the client s safety and comfort Thes...

Page 10: ...the Transfer Aid are operated by adjusting the leg lever plate as shown in the photo below To open the legs press on open Red Arrow and then to close the legs press on close Blue Arrow Transfer Aid ma...

Page 11: ...repositioning the Seat Pads 3 Using your hands grab the Seat Pads shown in Figure 12 and rotate them around until they reach the closed position Make sure the seat pads are position in the correct po...

Page 12: ...pport 6 While the patient is being lifted once possible attach the Safety Belt to the second Clip Pin on the Transfer Aid 7 Safety Belt operation complete To remove the sling simply reverse the proces...

Page 13: ...999562 Revision A P a g e 13 of 28 Technical Specification 5 1 TA400 Dimensions All dimensions below are shown in millimetres Figure 16 Figure 17 Figure 18 Figure 19...

Page 14: ...Foot Tray 5 5 kg 12 1 lb Table 5 Operational Forces Operating Force with 180 kg 400 lb 57 N 12 8 lbf Operating Force with 0 kg 0 lb 20 N 4 5 lbf Brakes 35 N 7 9 lbf Seat Pads 18 N 4 0 lbf Table 6 5 3...

Page 15: ...ycle and can no longer perform to its intended use safely the Aid must be decommissioned by an approved Service Engineer The following specifies the importance of correct disposal procedure including...

Page 16: ...nd If this does not correct the fault then contact your local authorised dealer immediately so the Transfer Aid can be checked to ensure proper continued operation General Inspection and Maintenance 9...

Page 17: ...Pads should align when in closed position Brakes Ensure the Brakes function when activated Ensure the Transfer Aid can move freely when the Brakes are deactivated Ensure they aren t rubbing on the Whe...

Page 18: ...onality of the device 5 minutes is a recommended contact time The device can withstand a longer contact period but the 5 minute recommendation as a minimum must be followed to provide an effective cle...

Page 19: ...turers Safety Data Sheet For a full review of the data please follow the link below http www nhsggc org uk media 236215 msds actichlor plus pdf Warranty This guarantee does not affect or in any way li...

Page 20: ...999562 Revision A P a g e 20 of 28 Service Record History Complete this section after each service repair inspection and or maintenance...

Page 21: ...999562 Revision A P a g e 21 of 28...

Page 22: ...999562 Revision A P a g e 22 of 28...

Page 23: ...999562 Revision A P a g e 23 of 28...

Page 24: ...999562 Revision A P a g e 24 of 28...

Page 25: ...999562 Revision A P a g e 25 of 28 User notes...

Page 26: ...999562 Revision A P a g e 26 of 28 User notes...

Page 27: ...999562 Revision A P a g e 27 of 28 User notes...

Page 28: ...or omissions Mackworth operates a policy of continuous improvement Specifications and other data are subject to change without notice Dealer service contact details Manufacturer contact details Prism...