996553
–
Revision A
P a g e
|
7 of 17
Hoist Installation
This section will cover the following guidelines on how to remove the safety components from the pre-installed track, how to
install the charger into the track and how to install the Hoist into the track.
This section will cover all variant types of the TX hoist range.
Ceiling Track Access
Prior to installing the Trolley into the pre-installed track, the track entrance must first be opened. A pre-installed track should
include an end stop, safety bolt and an end cap. If these parts are not installed, it is essential that they are once the Trolley is
installed.
Follow the guidance below to remove the ceiling track components. It is recommended that a step ladder is used to perform the
task safer and easier.
1.
Remove the end cap from the end of the track. (Figure 4)
2.
Using two 13mm spanners, loosen the safety bolt and remove the securing nut from the end. (Figure 5)
3.
Remove the safety bolt from the track and remove the large rubber bumper held within the track. (Figure 6)
4.
Using a 5mm Allen Key, release the two screws securing the end stop and slide the end stop out of the track. (Figure 7)
To refit the end stop, safety bolt and end cap, follow the steps above in reverse order noting the following points:
•
Ensure the end stop bumper is facing into the ceiling track.
•
Ensure all fasteners are secure and cannot come undone without the use of tools.
•
When fitting a charging dock, an end stop is not required at that end of the track.
(The images in this section all refer to a Track Type 1 installation, but the same process applies for all the remaining track types)
Figure 4
Figure 5
Figure 6
Figure 7