background image

 

IM213D-MC Issue 10 © 2017                                                                                                                                                                 Page 3 of 8 

2. On selecting the ‘Test Volume’ press the blue 
 

Mode

 button for 3 secs. 

  The meter will display ‘PURGE’ and ‘CALIBRATE’ 
  will also start to flash. 
 
3. Purge the system of air by running fluid through 
  the system. 
 
4. Once purged of air the calibration process can be 
  started by pressing the blue Mode button. 
  The unit will display 

RUN

 and the ’Test Volume’.  

 

 

E.g. 

RUN 100

 

 
5. Run the Test Volume through the meter until 
  stipulated volume has been reached (e.g. 100). 
 
6. Once this volume has been reached press the 
  blue 

Mode

 button to stop the test. 

  The unit will now compare the  ‘Measured Volume’ 
  to the ‘Test Volume’ and perform an ‘Auto 
  Calibration’ if the difference between the two 
  volumes are within ± 8% of each other. 
 

Note: 
If the difference between the two volumes is greater 
than ± 8% of each other, the unit will display one of 
the following messages.. 

• 

 

ERROR LOW 

• 

  ERROR HIGH 

 

if these messages are displayed please contact your 
Macnaught agent for advice. 

 

Returning to ‘Operation’ Mode

 

At any stage the unit can be returned to the 

Operation

’ mode by pressing the  red ‘Reset’ button 

for approx. 3 secs. 
Any program changes will automatically be saved. 

 

Maintenance Procedures. 

4. Remove the rotors (item 6) and inspect the 
  condition of each.  
  Also investigate if there is the presence of any 
  foreign material in the meter body, that may inhibit 
  the rotors  performance 

 

Reassembly

 

1.  Please  note,  the  design  of  the  rotor  and  shaft 
 

assembly  ensures  that  the  rotors  can  only  be 

 

re-installed  with  the  correct  orientation.  (i.e.  with 

 

the  magnets  being  in  close  proximity  to  the 

 

Electronic module). 

 

When  replacing  the  rotors  the    top  face  of  the 

 

rotors should be flush with the sealing face of the 

 

meter  body.  If  they  sit  higher  than  the  sealing 

 

face remove, turn over and replace. 

 

2.  Replace  the  rotors  (Item  6)  onto  the  shafts  at  90 
  degrees  to  each  other  (as  per  diagram  below)  
  and  check  their  operation  by  turning  either  of  the 
  rotors.   
  If  the  rotors  are  not  ‘

in  mesh’ 

correctly  or  do  not 

  move  freely, remove one of the rotors and replace  
  correctly at 90 degrees to the other rotor.  

Programming Instructions 

Disassembly

 

Ensure  that  the  fluid  supply  to  the meter  is  
disconnected,  and  the  line pressure is released  
before disassembly,  with  the  exception for repair or   
maintenance to the LCD or PCB where it is not 
necessity to isolate the meter from flow. Refer to the 
exploded parts diagram on subsequent pages for 
item numbers. 
 
1. Pull off protective boot (item 1) and unscrew the  
    four retaining screws (item 2) next remove the  
    Electronic Module (item 3) 

2. Check for evidence of moisture into the electronic 
  housing. If there is evidence of this, check the 
  condition of the O-Ring (item 4)  

3. Check the Rotors (item 6) rotate freely. 

4. Replace the 0-Ring (item 7) into the groove of the 
  Meter Cap (item 8). 

5. Replace the Meter Cap onto Meter Body (item 5) 
  Tighten Meter Cap screws (item 9) in a diagonal 
  sequence  
 

E.g. 1, 5, 3, 7, 4, 8, 6, 2 

6. Place the O-Ring (item 4) into the Electronic 
  Module (item 3) and mount the Electronic module 
  onto to the Meter Body 

7. Replace and tighten the Retaining Screws (item 2) 
  in diagonal sequence. 

8. Align and push on the protective boot (item 1) onto 
the electronic module (item 3). 

9. Before returning to service test the meter by 
  turning the Rotors with your finger. Or applying a 
  very low air pressure (no more than a good 
  breath) to the meter

 

3. To access the Rotor assembly, remove the 8 
  Meter Cap screws (item 9)  

Summary of Contents for IM012E

Page 1: ...design This has proven to be a reliable and highly accurate method of measuring flow Exceptional repeatability and high accuracy over a wide range of fluid viscosities and flow rates are features of the oval rotor design With a low pressure drop and high pressure rating oval rotor flow meters are suitable for both gravity and pump in line Before use confirm the fluid to be used is compatible with ...

Page 2: ...as shown above 4 Once the required Units have been selected move to the next section Calibration by pressing the Red Reset button Programming Instructions Note Any changes made during the programming phase will automatically be Saved when the unit is returned to the operation mode Accessing Programming Menu To enter in the programming Menu press the reset button for 5 secs Once in the programming ...

Page 3: ...ing the rotors the top face of the rotors should be flush with the sealing face of the meter body If they sit higher than the sealing face remove turn over and replace 2 Replace the rotors Item 6 onto the shafts at 90 degrees to each other as per diagram below and check their operation by turning either of the rotors If the rotors are not in mesh correctly or do not move freely remove one of the r...

Page 4: ...mbly procedure Follow step 1 to isolate the Electronic module 2 See Photograph below Remove the PCB from clear plastic housing by unscrewing the 3 retaining screws 3 The battery can now be removed by placing a screw driver into the slot slot indicated by arrow on the PCB and easing the battery free from its compartment 4 Replace with a new CR2450 Lithium battery Yellow Green Reed Switch Please Not...

Page 5: ...s Order for replacement Item no No off Part or Set Description 1 1 IM215s Protective boot 2 7 Screw M3 x 8 Pan Hd 3 1 DKIT IM012 MC Electronic module 4 1 O ring BS040 5 1 n a Body 6 2 MKIT IM012 01 Rotor 7 1 O ring BS035 8 1 n a Cover plate 9 8 n a Screw M4 x 16 Pan Hd ...

Page 6: ...IM213D MC Issue 10 2017 Page 6 of 8 Meter Dimensions Notes ...

Page 7: ...otors see instructions Meter not giving a pulse signal a Faulty hall effect sensor b Faulty reed switch c Magnets failed a Replace PCB Board b Replace PCB Board c Replacement rotor assembly required LCD register not working a Battery flat b Faulty LC Display a Replace battery b Replace PCB module WEEE Directive Waste Electrical and Electronic Equipment The WEEE Directive requires the recycling of ...

Page 8: ... Ethelene Glycol Based Anti Freeze Anti Boil mixture Max 50 w ater Wetted Materials Acetal Aluminium Steel Nitrile Rubber Connections 1 2 BSPT or 1 2 NPT Maximum Temperature 55 deg C 131 deg F Minimum Temperature 14 deg C 6 8 deg F Maximum Viscosity 1000cP Centipoise Re settable Batch Total 99999 9 Non Resettable Total 999999 Note This product should be disposed of according to all applicable loca...

Reviews: