1. 

Connect the RFC1000 wireless transceiver (sold separately) to the Windows PC with the 
provided USB cable.

2. 

Additional RFC1000’s can be used as repeaters to transmit over greater distances. If 
transmitting over a distance greater than  500 feet indoors, 2,000 feet outdoors or there are 
walls, obstacles or corners that need to be maneuvered around, set up additional RFC1000’s 
as needed. Plug each one into an electrical outlet in the desired locations.

3. 

Verify that the data loggers are in wireless transmission mode. Push and hold the 

Wireless

 

button on the data logger for 5 seconds to activate or deactivate wireless communication.

4. 

On the Windows PC, launch the MadgeTech 4 Software. 

5. 

All active data loggers will be listed in the Device tab within the Connected Devices panel.

6. 

To claim a data logger, select the desired data logger in the list and click the 

Claim

 icon.

7. 

Once the data logger has been claimed, select a start method in the Device tab.

Activating & Deploying the Data Logger

Different wireless channels may be used to create multiple networks in one area, or to avoid 
wireless interference from other devices. Any MadgeTech data logger or RFC1000 wireless 
transceiver that is on the same network is required to use the same channel. If all of the 
devices are not on the same channel, the devices will not communicate with one another. 
MadgeTech wireless data loggers and RFC1000 wireless transceivers are programmed by 
default on channel 25.

Changing the channel settings of the Elem

e

nt CO2 

1. 

Switch the wireless mode to 

OFF

 by holding down the 

Wireless

 button on the data logger for 5 

seconds. 

2. 

Using the provided USB cable, plug the data logger into the PC.

3. 

Open the MadgeTech 4 Software. Locate and select the data logger in the 

Connected Devices

 panel. 

4. 

In the Device tab, click the 

Properties

 icon.

5. 

Under the Wireless tab, select a desired channel (11 - 25) that will match with the RFC1000. 

6. 

Save all changes.

7. 

Disconnect the data logger.

8. 

Return the device to wireless mode by holding down the 

Wireless

 button for 5 seconds. 

Channel Programming

To configuring the channel settings of the RFC1000 wireless transceiver (sold separately), 
please refer to the RFC1000 Product User Guide that shipped with the product or download it 
from the MadgeTech website at 

www.madgetech.com

CHANNEL NOTE: MadgeTech wireless data loggers and wireless transceivers purchased prior to 

April 15, 2016

 are programmed by default to channel 11. Please refer to the Product User Guide 

provided with these devices for instructions to change the channel selection if needed.

Summary of Contents for Element CO2

Page 1: ...Product User Guide Element CO2 Wireless CO2 Humidity Temperature Data Logger ...

Page 2: ...at www madgetech com Quick Start Steps 3 Product Overview 4 Software Installation Guide 5 Activating Deploying the Data Logger 6 Channel Programming 6 Product Maintenance 7 RMA Instructions 7 Troubleshooting 8 Compliance Information 9 General Specifications 10 11 Table of Contents ...

Page 3: ...icking Start 7 To download data select the device in the list click the Stop icon and then click the Download icon A graph will automatically display the data 1 Install the MadgeTech 4 Software and USB Drivers onto a Windows PC 2 Confirm the data logger is not in wireless mode If wireless mode is on press and hold the Wireless button on the device for 5 seconds 3 Connect the data logger to the Win...

Page 4: ... information displayed on the LCD Screen Units Allows users to change displayed units of measurement to either Fahrenheit Celsius RH or ppm Wireless Push and hold this button for 5 seconds to activate or deactivate wireless communication Users have the ability to manually reset the statistics within the device to zero without needing to use the MadgeTech 4 Software Any data recorded up to that poi...

Page 5: ...cations from any internet enabled device Transmit real time data to the MadgeTech Cloud Services platform via the MadgeTech Data Logger Software running on a central PC or transmit directly to the MadgeTech Cloud without a PC using the MadgeTech RFC1000 Cloud Relay sold separately Sign up for a MadgeTech Cloud Services account at www madgetech com cloud Installing the USB Interface Driver USB Inte...

Page 6: ... transceiver that is on the same network is required to use the same channel If all of the devices are not on the same channel the devices will not communicate with one another MadgeTech wireless data loggers and RFC1000 wireless transceivers are programmed by default on channel 25 Changing the channel settings of the Element CO2 1 Switch the wireless mode to OFF by holding down the Wireless butto...

Page 7: ...l will be automatically sent when MadgeTech has received the RMA Product Maintenance Battery Replacement Materials U9VL J Battery 1 On the bottom of the data logger open the battery compartment by pulling in on the cover tab 2 Remove the battery by pulling it from the compartment 3 Install the new battery taking note of the polarity 4 Push the cover closed until it clicks Recalibration CO2 Zero Re...

Page 8: ...ooting wireless transceiver problems Check that the software properly recognizes the connected RFC1000 wireless transceiver If the wireless data logger is not appearing in the Connected Devices list it may be that the RFC1000 is not properly connected 1 In the MadgeTech 4 Software click the File button then click Options 2 In the Options window click Communications 3 The Detected Interfaces box wi...

Page 9: ...pt RSS standard s Operation is subject to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Le présent appareil est conforme aux CNR d Industrie Canada applicables aux appareils radio exempts de licence L exploitation est autorisée aux deux conditions suivantes...

Page 10: ... RFC1000 CE RFC1000 IP69K or RFC1000 Cloud Relay Transmission Distance Unobstructed line of sight outdoors 2 000 feet max Typical urban environment indoors 500 feet max Wireless BATTERY WARNING BATTERY MAY LEAK FLAME OR EXPLODE IF DISASSEMBLED SHORTED CHARGED CONNECTED TOGETHER MIXED WITH USED OR OTHER BATTERIES EXPOSED TO FIRE OR HIGH TEMPERATURE DISCARD USED BATTERY PROMPTLY KEEP OUT OF REACH OF...

Page 11: ...ireless Operation RFC1000 wireless transceiver PC Compatibility MadgeTech 4 Windows 10 8 7 Vista Window XP with SP3 MadgeTech Cloud Services Compatible Yes Operating Environment 0 C to 55 C 32 F to 131 F 0 RH to 95 RH non condensing 13 77 PSIA to 100 PSIA Dimensions Data logger 3 0 in x 3 5 in x 0 95 in 76 2 mm x 88 9 mm x 24 1 mm Sensor 1 98 in x 1 98 in x 1 15 in 50 3 mm x 50 3 mm x 29 2 mm Cabl...

Page 12: ...Base online at www madgetech com kbase Contact our friendly Customer Support Team at 603 456 2011 or support madgetech com Need Help MadgeTech 4 Software Support Refer to the built in help section of the MadgeTech 4 Software Download the MadgeTech 4 Software Manual at www madgetech com mt4 software manual pdf MadgeTech Cloud Services Support Download the MadgeTech Cloud Services Software Manual at...

Reviews: