Element CO2
7
RMA Instructions
To send a device back in to MadgeTech for calibration, service or repair, follow the instructions
below to create an RMA (Return Merchandise Authorization) on the MadgeTech website:
1.
Go to
www.madgetech.com/rma
.
2.
Sign into your account. For new customers, select
Create an Account
.
3.
Once signed in, click on the
Make New RMA
button.
4.
Complete the applicable fields on the form including customer Billing and Shipping
information, even if they are the same. Please see the field explanation below for a more
detailed description about questions asked in the Device Information section.
5.
When all of the fields are complete, click
Generate RMA
.
6.
Print out the confirmation page that follows containing the RMA number and MadgeTech’s
address for shipping.
A Return Merchandise Authorization must be accompanied by a
copy of the RMA paperwork and shipping is prepaid by the customer
. The RMA number
should be clearly marked on the outside of the package.
7.
Ship the package via UPS, FedEx, TNT, or DHL to the address listed on the confirmation
page.
USPS will not ship MadgeTech data loggers
.
8.
A notification email will be automatically sent when MadgeTech has received the RMA.
Product Maintenance
Battery Replacement
Materials:
U9VL-J Battery
1.
On the bottom of the data logger, open the battery compartment by pulling in on the cover tab.
2.
Remove the battery by pulling it from the compartment.
3.
Install the new battery, taking note of the polarity.
4.
Push the cover closed until it clicks.
Recalibration
CO
2
Zero Reset:
The Elem
e
nt CO2 is equipped with a self-calibrating sensor that can be reset by the user at
any time using the CO
2
Zero Reset feature. The reset feature adjusts the calibration of the CO
2
channel by taking the current reading of the sensor and “zeroing” it to 400 ppm (outside air
concentration). To do this expose the data logger to outside fresh air for at least five minutes to
stabilize. Then press and hold the scroll and wireless keys for three seconds at the same time.
The message “zeroing CO
2
” will appear on the device screen. The message will disappear within
30 seconds and the device will beep twice confirming the zero operation is complete.
Depending upon device exposure to extreme parameters and frequency of use, MadgeTech
suggests performing the CO
2
zero reset as frequent as needed for application accuracy.
CO
2
channel calibrated accuracy: ± 70ppm + 5% of range, add 8ppm per day of continuous
monitoring from zero reset.