 Product User Guide  |  3

HiTemp140-FP

PRODUCT USER GUIDE

Device Operation (cont’d)

Trigger Settings

The device can be programmed to only record based off 
user configured trigger settings.
1.  In the 

Connected devices

 panel, select the intended 

device to change the settings.

2.  On the 

Device

 tab, in the 

Information

 group, click 

Properties

. Users can also right-click on the device 

and select 

Properties

 in the context menu.

3.  Click 

Trigger

 and configure the Trigger settings.  

Trigger formats are available in Window and Two Point 
(bi-level) mode. Window mode allows for one range of 
temperature monitoring and two point mode allows 
for two ranges.

Set Password

To password protect the device so that others cannot 
start, stop or reset the device:
1.  In the 

Connected Devices

 panel, click the device 

desired.

2.  On the 

Device

 Tab, in the 

Information

 Group, click 

Properties

. Or, right-click the device and select 

Properties

 in the context menu.

3.  On the 

General

 Tab, click 

Set Password

.

4.  Enter and confirm the password in the box that 

appears, then select 

OK

.

Device Maintenance

O-Rings

O-ring maintenance is a key factor when properly 
caring for the HiTemp140-FP. The O-rings ensure a tight 
seal and prevent liquid from entering the inside of the 
device. Please refer to the application note 

O-Rings 101: 

Protecting Your Data

, found a

madgetech.com

, for 

information on how to prevent O-ring failure.

Battery Replacement

Materials:

 ER14250MR-145 Battery

1.  Unscrew the bottom of the 

logger and remove the battery.

2.  Place the new battery into the 

logger. Note the polarity of 
the battery. It is important to 
insert the battery with positive 
polarity pointing upward 
towards the probe. Failure to 
do so could result in product inoperability or potential 
explosion if exposed to high temperatures.

3.  Screw the cover back onto the logger.

Recalibration 

MadgeTech recommends annual recalibration. To send 
devices back for calibration, visit 

madgetech.com

.

Notice: Steam Sterilization Applications

The pervasive nature of pressurized steam creates a very 
difficult environment for electronics. Please refer to the 
following preventative maintenance procedure when 
using this device in steam sterilization applications. 
Additionally, this device is not ideal for steam sterilization 
applications above 121 °C/1.1 bar.

Preventive Maintenance

After every 3 hours of steam exposure:
1.  Remove the endcap and battery from the device (ref. 

battery change procedure on Product User Guide)

2.  Place open logger (minus battery) in an oven at 120 °C 

(250 °F) for a minimum of 30 minutes

3.  Remove logger from oven and allow to cool to room 

temperature

4.  Re-assemble the logger with the battery (note polarity) 

and endcap

Note:

 This product is rated for use up to 140 °C  

(284 ºF). Please heed the battery warning. The 

product will explode if exposed to temperatures 

above 140 °C (284 ºF).