background image

 

Product Notes

Getting Started

The HiTemp150FP-TSK consists of a HiTemp150FP data logger in a thermal enclosure. The 
thermal enclosure allows the device to withstand higher temperatures for certain durations 
of time. The chart located below outlines the time versus temperature durations.

To start the logger, unscrew the thermal enclosure and separate the thermal enclosure lid, 
from the body of the data logger. Unscrew the knurled nut to access the communication 
port. Screw the knurled nut back on.

The logger must be removed from the barrier immediately after removal from the heat 
environment. Be careful, the logger may be very hot.

Operating Environment

The HiTemp150FP-TSK has a time limited operating environment of between -160 °C to 
+250 °C, and 0 %RH to 95 %RH, non-condesing. 

O-Rings

O-ring maintenance is a key factor when properly caring for the HiTemp150FP-TSK. The 
o-rings ensure a tight seal and prevent liquid from entering the inside of the device.
Please refer to the application note “O-Rings 101: Protecting Your Data”, found on the 
MadgeTech website, for information on how to prevent O-ring failure.

Installation Guide

Installing the Interface cable

- IFC200 

Insert the device into a USB port. The drivers will install automatically.

- IFC110 

Plug the serial cable into the port and verify it is secure.

- USB-1 or USB-101 

Install the USB drivers from the USB software stick provided in the kit, then plug the USB 
cable into the computer and the serial cable into the serial port.

Product Information Card

HiTemp150FP-TSK

Installing the software

Insert the Software USB into an open port. If the autorun does not appear, locate the drive 
on the computer and double click on 

Autorun.exe

. Follow the instructions provided in the 

Wizard.

Device Operation

Connecting and Starting the data logger

- Once the software is installed and running, plug the interface cable into the data logger.
- Connect the USB end of the interface cable into an open USB port on the computer. 
- The device will appear in the Connected Devices list, highlight the desired data logger.  
- For most applications, select “

Custom Start

” from the menu bar and choose the desired 

start method, reading rate and other parameters appropriate for the data logging 
application and click “

Start

”. (

Quick Start

” applies the most recent custom start options, 

Batch Start

” is used for managing multiple loggers at once, “

Real Time Start

” stores the 

dataset as it records while connected to the logger.

)

- The status of the device will change to “

Running

”, “

Waiting to Start

” or “

Waiting to 

Manual Start

”, depending upon your start method.

- Disconnect the data logger from the interface cable and place it in the environment to 

measure.

 

Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device 
cannot be restarted until it has been re-armed by the computer.

 

Downloading data from a data logger

- Connect the data logger to the interface cable.
- Highlight the data logger in the Connected Devices list. Click “

Stop

” on the menu bar.

- Once the data logger is stopped, with the logger highlighted, click “

Download

”. You will 

be prompted to name your report.

- Downloading will offload and save all the recorded data to the PC.

Device Maintenance

Battery Replacement

The HiTemp150FP-TSK must be sent to MadgeTech for battery replacement.

Recalibration 

The HiTemp150FP standard calibration is two points at 50 °C and 150 °C. 

Pricing:

Recalibration traceable to NIST 

$70.00

Recalibration $40.00
 

Prices and specifications subject to change. See MadgeTech’s terms and conditions at www.madgetech.com.
To send the devices back, visit www.madgetech.com, select Services then RMA Process.

Ambient Temperature

Time in Air to Max Internal Temp

(150 °C/302 °F)

160 °C (320 °F)

75 minutes

170 °C (338 °F)

63 minutes

180 °C (356 °F)

55 minutes

190 °C (374 °F)

50 minutes

200 °C (392 °F)

45 minutes

210 °C (410 °F)

42 minutes

220 °C (428 °F)

39 minutes

230 °C (446 °F)

36 minutes

240 °C (464 °F)

34 minutes

250 °C (482 °F)

32 minutes

Reviews: