Troubleshooting Tips

Why is the wireless data logger not appearing in the software?

If the RFRHTemp2000A doesn’t appear in the Connected Devices panel, or an error message is 
received while using the RFRHTemp2000A, try the following: 

• Check that the RFC1000 is properly connected. For more information, see Troubleshooting Interface

Cable problems (below).

• Ensure that the battery is not discharged. For best voltage accuracy, use a voltage meter connected to

the battery of the device. If possible, try switching the battery with a new 9V lithium.

• Ensure that no other MadgeTech software (such as 

MadgeTech 2,

 or 

MadgeNET

) is running in the

background.

• Ensure that 

MadgeTech 4

 is being used. 

MadgeTech 2

 and 

MadgeNET

 are not the correct software for

connecting with the RFRHTemp2000A.

• Ensure that the 

Connected Devices 

panel is large enough to display devices. This can be verified by

positioning the cursor on the edge of the Connected Devices panel until the resize cursor appears, then
dragging the edge of the panel to resize it. The screen layout may also be reset in the options menu by
selecting 

File

Options

, and scrolling to the bottom.

Troubleshooting Interface Cable problems

Check that the software properly recognizes the connected RFC1000 wireless transceiver.

If the wireless data logger is not appearing in the 

Connected Devices

 list, it may be that the RFC1000 is

not properly connected.

1.

In the software, click the 

File

 button, then click 

Options

.

2.

In the 

Options

 window, click 

Communications

.

3.

The 

Detected Interfaces

 box will list all of the available communication interfaces. If the RFC1000 is

listed here, then the software has correctly recognized and is ready to use it.

Check that Windows recognizes the connected RFC1000 wireless transceiver.

If the software does not recognize the RFC1000, there may be a problem with Windows or the USB drivers.

1.

In Windows, click 

Start

, right-click 

Computer

 and choose 

Properties

 or press 

Break

 as a

keyboard shortcut.

2.

Click 

Device Manager

 in the left hand column.

3.

Double click 

Universal Serial Bus Controllers

.

4.

Look for an entry for

 Data logger Interface

.

5.

If the entry is present, and there are no warning messages or icons, then windows has correctly

recognized the connected RFC1000.

6.

If the entry is not present, or has an exclamation point icon next to it, the USB drivers may need to

be installed. These are available on the software flash drive included with the RFC1000, and on the
MadgeTech website.

Ensure that the USB end of the RFC1000 is securely connected to the computer.
1.

Locate the USB-A plug of the RFC1000.

2.

If the interface cable is connected to the PC, unplug it. Wait ten seconds.

3.

Reconnect the cable to the PC.

4.

Check to make sure that the red LED is lit, indicating a successful connection.

DOC 1268036-00  REV 6  2015.01.02

Summary of Contents for MadgeNet RFRHTemp2000A

Page 1: ...RFRHTemp2000A RFRHTemp2000A Wireless Temperature and Humidity Data Logger Product User Guide...

Page 2: ...ructions 4 Product Maintenance 5 Recalibration 5 RMA Instructions 5 General Specifications 6 7 RFC1000 Transceiver Product Overview 9 Installation Guide 9 Mounting Instruction 10 Channel Programming 1...

Page 3: ...e Start icon 7 To download data select the device in the list click the Stop icon and then click the Download icon A graph will automatically display the data Quick Start Steps 1 Install the MadgeTech...

Page 4: ...er for mounting purposes Selection Buttons The logger is designed with 3 direct selection buttons Scroll Units and Wireless Scroll Allows user to scroll through current readings average stats and devi...

Page 5: ...will occur from the drop down menus 4 If selecting to receive notifications by email or text message click Email settings and make sure the correct info has been entered Email and text message notifi...

Page 6: ...urely snap to the backside of the logger for wall mounting if desired as shown below There are two holes in the base to allow for screws for mounting Mounting Instructions 4 Still need help For more t...

Page 7: ...Recalibration is recommended annually for any MadgeTech data logger a reminder is automatically displayed in the software when the device is due Countries approved for use purchase and distribution o...

Page 8: ...6 2 mm x 88 9 mm x 24 1 mm Data logger only Weight 4 1 oz 0 2563 lbs Enclosure Material ABS Plastic Alarm User configurable audible on screen email and text SMS alarms Alarm Delay A cumulative alarm d...

Page 9: ...ransmission Distance to other RFC1000 s RFC1000 4 000 ft max outdoors line of sight unobstructed 1 000 ft max indoors typical urban environment RFC1000 CE 2 500 ft max outdoors line of sight unobstruc...

Page 10: ...Product User Guide RFC1000 Wireless Transceiver 8 Other models available RFC1000 CE Approved for European markets RFC1000 IP69K Splash Proof for harsh environments...

Page 11: ...0 C to 85 C and a humidity range of 0 RH to 95 RH non condensing The RFC1000 is rated IP40 and is protected against solids that are greater than 1mm in size This device is not water resistant LEDs The...

Page 12: ...ss performance both the RFC1000 and the MadgeTech data loggers should be mounted in the same orientation This usually means that the external antenna should be pointing straight up The antenna can piv...

Page 13: ...dgeTech data logger or RFC1000 that is on the same network is required to use the same channel If all of the devices are not on the same channel the devices will not communicate with one another The i...

Page 14: ...there are walls obstacles corners that need to be maneuvered around set up additional RFC1000s as needed Step 3 Verify that the data loggers are in wireless transmission mode See Channel Programming...

Page 15: ...ada applicables aux appareils radio exempts de licence L exploitation est autoris e aux deux conditions suivantes 1 l appareil ne doit pas produire de brouillage et 2 l utilisateur de l appareil doit...

Page 16: ...earing in the Connected Devices list it may be that the RFC1000 is not properly connected 1 In the software click the File button then click Options 2 In the Options window click Communications 3 The...

Reviews: