RFRHTemp2000A
Product Maintenance
Battery Replacement
Materials:
U9VL-J Battery
1.
On the bottom of the enclosure, open the battery compartment by pulling in on the
cover tab.
2.
Remove the battery by pulling it from the compartment.
3.
Install the new battery, taking note of the polarity.
4.
Push the cover closed until it clicks.
Recalibration
Standard recalibration for the RFRHTemp2000A is one point at 25 °C for the internal
temperature channel, and two points at 25 %RH and 75 %RH for the humidity channel.
Recalibration is recommended annually for any MadgeTech data logger; a reminder is
automatically displayed in the software when the device is due.