RFTCTemp2000A
Product Maintenance
Battery Replacement
Materials:
U9VL-J Battery
1.
On the bottom of the enclosure, open the battery compartment by pulling in on the
cover tab.
2.
Remove the battery by pulling it from the compartment.
3.
Install the new battery, taking note of the polarity.
4.
Push the cover closed until it clicks.
Recalibration
Standard recalibration for the RFTCTemp2000A is two points, 25 °C and 60 °C for the
internal channel (ambient), and 0 mV for the thermocouple (remote) channel.
Recalibration is recommended annually for any MadgeTech data logger; a reminder is
automatically displayed in the software when the device is due.