Product User Guide | 3
PR140
PRODUCT USER GUIDE
Device Operation (cont’d)
Set Password
To password protect the device so that others cannot
start, stop or reset the device:
1. In the
Connected Devices
panel, click the device
desired.
2. On the
Device
Tab, in the
Information
Group, click
Properties
. Or, right-click the device and select
Properties
in the context menu.
3. On the
General
Tab, click
Set Password
.
4. Enter and confirm the password in the box that
appears, then select
OK
.
Device Maintenance
O-Rings
O-ring maintenance is a key factor
when properly caring for the PR140.
The O-rings ensure a tight seal and
prevent liquid from entering the
inside of the device. Please refer to
the application note
O-Rings 101:
Protecting Your Data
, found at
, for information
on how to prevent O-ring failure.
Battery Replacement
Materials:
ER14250MR-145 Battery
1. Unscrew the bottom of the logger and remove the
battery.
2. Place the new battery into the logger. Note the polarity
of the battery. It is important to insert the battery with
positive polarity pointing upward towards the probe.
Failure to do so could result in product inoperability or
potential explosion if exposed to high temperatures.
3. Screw the cover back onto the logger.
Recalibration
MadgeTech recommends annual recalibration. To send
devices back for calibration, visit
.
Note:
This product is rated for use up to 140 °C
(284 ºF). Please heed the battery warning. The
product will explode if exposed to temperatures
above 140 °C (284 ºF).