background image

Product User Guide

7

Front Panel Overview

Changing the display units

The PR2000 comes with factory default display units of PSI for pressure and the real-time 
pressure graphing feature. These units can be easily changed by pressing the F3 button 
in the main screen and then selecting F1 for pressure or F2 for the pressure graph. After 
selecting the channel, the available units can be scrolled through by either pressing the 
channel’s function key repeatedly or using the UP and DOWN keys

 

Button pressing chain:

 Main Screen  -> F3 -> F1 (pressure) or F2 (pressure graph) ->  

function key repeatedly or UP and DOWN

Checking the memory status

A status icon appears on all screens representing memory, but further information includ-
ing percent memory left and number of readings taken can also be viewed. From the Main 
Screen press the F1 key to enter the Status screens then press F2 to view memory status 
information.

Button pressing chain:

 

Main Screen -> F1 -> F2

 

Checking power status

A battery status and external power status (if available) icon appear on all screens, but per-
cent battery power remaining and external power presence as well as battery type, current 
battery voltage, and current external voltage can also be viewed. From the Main Screen 
press F4 to view the Device Configuration Menu, F2 to access the power options, then F4 
twice to view the Power Status screen, including battery power percent remaining and the 
presence of external power.  Battery type and battery voltage are also displayed, as well as 
external power voltage (if connected).
 

Button pressing chain:

 Main Screen -> F4 -> F2 -> F4 -> F4 

 

Changing the Contrast

The PR2000’s LCD screen contrast values can be changed in two ways. One method is 
outlined in the Function Reference Guide. A faster, simpler way involves simultaneously 
pressing the CANCEL and UP or DOWN button in any screen.

Button pressing chain:

 

UP (to increase) or DOWN (to decrease)

Summary of Contents for PR2009

Page 1: ...PR2000 PR2000 Pressure Data Logger with LCD Product User Guide...

Page 2: ...dgeTech product line visit our website at madgetech com Quick Start Steps 3 Product Overview 4 Software Installation 5 Device Operation 5 Computer Interface 6 Front Panel Overview 7 Screen Description...

Page 3: ...d separately 4 Launch the MadgeTech 4 Software The pHTemp2000 will appear in the Connected Devices window indicating the device has been recognized 5 Select the start method reading rate and any other...

Page 4: ...age statistics A trending graph is also displayed of the last 100 readings Please refer to the PR2000 Product Manual for full product details by going to madgetech com product documentation LED Indica...

Page 5: ...red data logger For most applications select Custom Start from the menu bar and choose the desired start method reading rate and other parameters appropriate for the data logging application and click...

Page 6: ...nto the female receptacle of the data logger Fully insert the female USB connector into the USB Please see the Data Logger Software manual for further information WARNING Install driver before connect...

Page 7: ...s screens then press F2 to view memory status information Button pressing chain Main Screen F1 F2 Checking power status A battery status and external power status if available icon appear on all scree...

Page 8: ...ns Run Parameters Statistics Menu Screen Displays options available within the statistics menu Type Statistics Displays statistics from the pressure channel Memory Status Channel Statistics Displays s...

Page 9: ...ns Hardware and Power Interruption Power Interruption Displayed as notification when power is interrupted during device operation F1 OK accepts notification and displays main screen F2 no function F3...

Page 10: ...ve the battery from its compartment and unsnap it from the connector Snap the new battery into the terminals and verify it is secure Replace the cover taking care not to pinch the wires Screw the encl...

Page 11: ...ery Life 1 year battery life at 1 minute reading rate with display off 30 days typical with continuous display use Operating Environment 20 C to 60 C 4 F to 140 F 0 RH to 95 RH non condensing Material...

Page 12: ...MadgeTech Inc 6 Warner Road l Warner NH 03278 Phone 603 456 2011 l Fax 603 456 2012 madgetech com l info madgetech com DOC 1148036 00 REV 15 2019 07 16...

Reviews: