Product Notes
The PRHTemp101 and PRHTemp110 both measure ambient temperature, humidity and
barometric pressure. The temperature and humidity sensors are internal to the device and
the port for the pressure sensor is located on the front of the device.
LEDs
Once started, the LED will flash at the selected reading rate to indicate that the device is
running.
Device Maintenance
Battery Replacement
Materials
:
Small Phillips Head Screwdriver
Replacement Battery (LTC-7PN)
- Puncture the center of the back label with the screw driver and unscrew the enclosure.
- Remove the battery by pulling it perpendicular to the circuit board.
- Insert the new battery into the terminals and verify it is secure.
- Screw the enclosure back together securely.
Note: Be sure not to over tighten the screws or strip the threads.
Recalibration
The PRHTemp101 or PRHTemp110 standard calibration is performed at one temperature
point at 25°C, two humidity points at 25%RH and 75%RH and two pressure points at 14.7
PSI and 30 PSI.
Pricing:
Recalibration traceable to NIST
$120.00
Recalibration
$100.00
Additional Services
:
As Found Data
$15.00 per parameter/channel
Verification Point
$15.00 per point
Prices and specifications subject to change. See MadgeTech’s terms and conditions at www.madgetech.com.
To send the devices back, visit www.madgetech.com, select Services then RMA Process.
PRHTemp101 and PRHTemp110
Product Quick Reference Card
Installation Guide
Installing the Interface cable
- IFC200
Insert the device into a USB port. The drivers will install automatically.
- IFC110
Plug the serial cable into the port and verify it is secure.
Installing the software
Insert the Software CD in the CD-ROM Drive. If the autorun does not appear, locate the
drive on the computer and double click on
Autorun.exe
. Follow the instructions provided
in the Wizard.
Connecting the data logger
- Once the software is installed and running, plug the interface cable into the data logger.
- Click the
Communication Menu
, then
Auto Configure Port
.
- After a moment, a box will appear stating a device has been found.
- Click
OK
. The
Device Status
box will appear. Click
OK
.
- At this point, communications have been configured for your logger. These settings can
be found under the
Communication Menu
.
Note: For additional installation instructions refer to your “Data Logger & Software Operating Manual”.
Device Operation
Starting the data logger
- Click
Device Menu
then
Start Device
.
- Choose the desired start method.
- Choose the start parameters by selecting a
Reading Rate
suitable for your application.
- Enter in any other desired parameters and click
Start
.
- A box will appear stating the data logger has been started. Click
OK
.
- Disconnect the data logger from the interface cable and place it in the environment to
measure.
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device
cannot be restarted until it has been re-armed by the computer.
Downloading data from a data logger
- Connect the data logger to the interface cable.
- Click the
Device Menu
then
Read Device Data
. This will offload all recorded data onto
the PC.