background image

Wiring the Data Logger

Wiring Options

For 4-wire RTD probes, connect the four lead wires to your RTD logger as shown in the 
figures below.

For 3-wire RTD probes, short inputs 3 
and 4 together, then connect the lead 
wires to inputs 1, 2 and 3.
For 2-wire RTD probes, short inputs 
3 and 4 together and inputs 1 and 2 
together, then connect the RTD lead 
wires to inputs 2 and 3.

 
Warning: Note the polarity instructions. Do not attach wires to the wrong terminals. 100 Ohm, 2 or 4 wire RTD probes are 
recommended for the most accurate performance. Most 100 Ohm, 3 wire RTD probes will work, but MadgeTech cannot guarantee 
the accuracy. To determine whether or not the 3-wire RTD probe will work, the resistance between the two same colored wires 
should be less than 1 Ohm. (Note: Please contact the manufacturer of the RTD probe for questions on the resistance)

 

 

Device Maintenance

Battery Replacement

Materials

:

3/32” HEX Driver (Allen Key)
Replacement Battery (U9VL-J)

- Remove the cover from the device by unscrewing the four screws.
- Remove the battery from its compartment and unsnap it from the connector.
- Snap the new battery into the terminals and verify it is secure.
- Replace the cover taking care not to pinch the wires. Screw the enclosure back together 

securely.

 
Note: Be sure not to over tighten the screws or strip the threads.

 
Recalibration 

The QuadRTD or OctRTD standard calibration is at two points, 50

 and 150

.

 
Pricing:

 

QuadRTD  OctRTD

Recalibration traceable to NIST 

$79.00 

$91.00

Recalibration 

$49.00 $61.00 

 

 
Additional Services:

Channel (1st) 

$45.00 at custom point

Additional Channels 

$4.50 at custom point

Verification Point 

$15.00 per point

 
Prices and specifications subject to change. See MadgeTech’s terms and conditions at www.madgetech.com

To send the devices back, visit www.madgetech.com, select Services then RMA Process. 

QuadRTD and OctRTD

Product Quick Reference Card

Installation Guide

Installing the Interface cable

- IFC200
  Insert the device into a USB port. The drivers will install automatically.
- IFC110
  Plug the serial cable into the port and verify it is secure.
- USB-1 or USB-101
  Install the USB drivers from the CD provided in the kit, then plug the USB cable into the 

computer and the serial cable into the serial port.

Installing the software

Insert the Software CD in the CD-ROM Drive. If the autorun does not appear, locate the 
drive on the computer and double click on 

Autorun.exe

. Follow the instructions provided 

in the Wizard.

Connecting the data logger

- Once the software is installed and running, plug the interface cable into the data logger. 
- Click the 

Communication Menu

, then 

Auto Configure Port

.

- After a moment, a box will appear stating that a device has been found. 
- Click 

OK

. The 

Device Status

 box will appear. Click 

OK

.

- At this point, communications have been configured for your logger. These settings can 

be found under the 

Communication Menu

.

 

Note: For additional installation instructions refer to your “Data Logger & Software Operating Manual”.

Device Operation

Starting the data logger

- Click 

Device Menu

 then

 Start Device

- Choose the desired start method.
- Choose the start parameters by selecting a 

Reading Rate

 suitable for your application.

- Enter in any other desired parameters and click 

Start

- A box will appear stating the data logger has been started. Click 

OK

- Disconnect the data logger from the interface cable and place it in the environment to 

measure.

 
Note: The device will stop recording data when the end of memory is reached or the device is stopped. At this point the device 
cannot be restarted until it has been re-armed by the computer. 

Downloading data from a data logger

- Connect the data logger to the interface cable.
- Click the 

Device Menu

 then 

Read Device Data

. This will offload all recorded data onto 

the PC. 

3-Red

4-Red or Short to 3

1-Black or Short to 2

2-Black

Connector for 100
Ohm PtRTD

Connector for interface cable

3

4

1

2

3

4

1

2

KEY

4- Excitation Current Out (+)

3- Measurement (+) Input

2- Measurement(-) Input

1- Ground (-)

Reviews: