Troubleshooting Tips

Why is the data logger not appearing in the software?

If the QuadTemp or OctTemp doesn’t appear in the Connected Devices panel, or an error message
is received while using either device, try the following: 

• Check that the IFC200 is properly connected. For more information, see Troubleshooting Interface

Cable problems (below).

• Ensure that the battery is not discharged. For best voltage accuracy, use a voltage meter connected to

the battery of the device. If possible, try switching the battery with a new 9V lithium. 

• Ensure that no other MadgeTech software (such as 

MadgeTech 2

) is running in the background.

• Ensure that the 

Connected Devices 

panel is large enough to display devices. This can be verified by

positioning the cursor on the edge of the Connected Devices panel until the resize cursor appears, then
dragging the edge of the panel to resize it. The screen layout may also be reset in the options menu by
selecting 

File

Options

, and scrolling to the bottom.

Troubleshooting Interface Cable problems 

Check that the software properly recognizes the connected IFC200 interface cable.

If either data logger is not appearing in the 

Connected Devices

 list, it may be that the IFC200 is not

properly connected.

1. 

In the software, click the 

File

 button, then click 

Options

2. 

In the 

Options

 window, click 

Communications

3. 

The 

Detected Interfaces

 box will list all of the available communication interfaces. If the IFC200 is

listed here, then the software has correctly recognized and is ready to use it.

Check that Windows recognizes the connected IFC200 interface cable.

If the software does not recognize the IFC200, there may be a problem with Windows or the USB drivers.

1. 

In Windows, click 

Start

, right-click 

Computer

 and choose 

Properties

 or press 

Break

 as a

keyboard shortcut. 

2. 

Click 

Device Manager

 in the left hand column.

3. 

Double click 

Universal Serial Bus Controllers

4. 

Look for an entry for

 Data logger Interface

.

5.

 If the entry is present, and there are no warning messages or icons, then windows has correctly 

recognized the connected IFC200. 

6.

 If the entry is not present, or has an exclamation point icon next to it, the USB drivers may need to

be installed. These are available on the software flash drive included with the IFC200, and on the
MadgeTech website.

Ensure that the USB end of the IFC200 is securely connected to the computer.
1. 

Locate the USB-A plug of the IFC200. 

2. 

If the interface cable is connected to the PC, unplug it. Wait ten seconds. 

3.

 Reconnect the cable to the PC.

4.

 Check to make sure that the red LED is lit, indicating a successful connection.

Advanced Technical Services Ltd: Ph +64 3 3774793 Fax +64 3 3774794

77 Brisbane Street Christchurch New Zealand

Summary of Contents for QuadTemp

Page 1: ...rature Data Logger OctTemp 8 Channel Thermocouple Based Temperature Data Logger Product User Guide Thermocouple Plugs Probes Sold Separately Thermocouple Plugs Probes Sold Separately To view the full...

Page 2: ...Steps 1 Product Overview 2 Additional Features Operation 3 4 Product Maintenance 5 Recalibration 5 RMA Instructions 5 General Specifications 6 9 Support Back Cover Table of Contents To view the full...

Page 3: ...r the data logging applica tion and click Start Quick Start applies the most recent custom start options Batch Start is used for managing multiple loggers at once Real Time Start stores the dataset as...

Page 4: ...stomers should choose the QuadTemp EU model under the ordering information This model includes a wall mounted 230 VAC 50 Hz 9 VDC power adapter The OctTemp is an eight channel thermocouple data logger...

Page 5: ...ces within the software Enable Disable Channels By default all channels are enabled 1 In the Connected Devices panel click the device desired 2 On the Device Tab in the Information Group click Propert...

Page 6: ...ming Up to a 10 character channel name can be programmed into the data logger for each channel This ability helps to rename a channel in a report to distinguish it from other similarly named channels...

Page 7: ...t RMA Process 2 When the web page opens please sign in If this is the first time select Click here to register an account and create one Once signed in click on the Make New RMA button and fill in all...

Page 8: ...5 C T 270 to 400 0 1 C 0 5 C E 270 to 980 0 1 C 0 5 C R 50 to 1760 0 5 C 2 0 C S 50 to 1760 0 5 C 2 0 C B 50 to 1820 0 5 C 2 0 C N 270 to 1300 0 1 C 0 5 C Start Modes Software programmable immediate...

Page 9: ...RH to 95 RH non condensing Dimensions 7 24 in x 2 7 in x 1 02 in 183 mm x 68 mm x 26 mm Material Black anodized aluminum Weight 15 2 oz 430 g Approvals CE QuadTemp General Specifications Cont 7 Batte...

Page 10: ...0 1 C 0 5 C K 270 to 1370 0 1 C 0 5 C T 270 to 400 0 1 C 0 5 C E 270 to 980 0 1 C 0 5 C R 50 to 1760 0 5 C 2 0 C S 50 to 1760 0 5 C 2 0 C B 50 to 1820 0 5 C 2 0 C N 270 to 1300 0 1 C 0 5 C Start Modes...

Page 11: ...Environment 20 C to 60 C 4 F to 140 F 0 RH to 95 RH non condensing Dimensions 7 24 in x 2 7 in x 1 26 in 183 mm x 68 mm x 32 mm Material Black anodized aluminum Weight 17 3 oz 490 g Approvals CE Batte...

Page 12: ...he software click the File button then click Options 2 In the Options window click Communications 3 The Detected Interfaces box will list all of the available communication interfaces If the IFC200 is...