background image

PRODUCT USER GUIDE

RTDTemp101A

DOC-1216035-00  |  REV 11  2020.05.27

NEED HELP?

Product Support & Troubleshooting:

• Visit our Resources online at 

madgetech.com/resources

.

• Contact our friendly Customer Support Team at 

01929 459 459 or support@wpls.co.uk

MadgeTech 4 Software Support:

• Refer to the built-in help section of the MadgeTech 4 Software.
• Download the MadgeTech 4 Software Manual at 

madgetech.com

.

• Contact our friendly Customer Support Team at 

01929 459 459 or support@wpls.co.uk

LED Indicators

Green LED blinks:

 10 seconds to indicate 

logging and 15 seconds to indicate delay 
start mode.

Red LED blinks:

 

10 seconds to indicate 

low battery and/or memory and 1 second 
to indicate an alarm condition. 

Multiple Start/Stop Mode Activation

•  

To start device:

 Press and hold the pushbutton for

5 seconds, the green LED will flash during this time. The
device has started logging.

•  

To stop the device:

 Press and hold the pushbutton for

5 seconds, the red LED will flash during the this time.
The device has stopped logging.

Device Maintenance

Battery Replacement

Materials: Small Phillips Head Screwdriver and a 
Replacement Battery (LTC-7PN)

1. Puncture the center of the back label with the screw

driver and unscrew the enclosure.

2. Remove the battery by pulling it perpendicular to the

circuit board.

3. Insert the new battery into the terminals and verify it

is secure.

4. Screw the enclosure back together securely.

Note:

 Be sure not to over tighten the screws or strip the threads.

Recalibration

Recalibration is recommended annually. To send 
devices back for calibration, visit 

support@wpls.co.uk or 

call 01929 459 459. 