Set Password

To password protect the device so that others cannot 
start, stop or reset the device:
1.  In the 

Connected Devices

 panel, click the device 

desired.

2.  On the 

Device

 Tab, in the 

Information

 Group, click 

Properties

. Or, right-click the device and select 

Properties

 in the context menu.

3.  On the 

General

 Tab, click 

Set Password

.

4.  Enter and confirm the password in the box that 

appears, then select 

OK

.

LED Functionality

Green LED:

  

Flashes 10 seconds to indicate logging and  
15 seconds to indicate delay start mode

Red LED:

  

Flashes 10 seconds to indicate low battery 
and/or memory and 1 second to indicate an 
alarm condition

Device Maintenance

Battery Replacement
Materials: 

Small Phillips Head Screwdriver and 

Replacement Battery (LTC-7PN)

1.  Open the ThermoVault and remove the data logger.
2.  Puncture the center of the back label with the screw 

driver and unscrew the enclosure.

3.  Remove the battery by pulling it perpendicular to the 

circuit board.

4.  Insert the new battery into the terminals and verify it is 

secure.

5.  Screw the enclosure back together securely.

Note:

 Be sure not to over tighten the screws or strip the threads.

Recalibration

Recalibration is recommended annually. To send devices 
back for calibration, visit 

madgetech.com

.

Thermocouple Type

To change the thermocouple type:
•  In the 

Connected Devices

 panel, click the device desired.

•  On the 

Device

 Tab, in the 

Information

 Group, click 

Properties

. Or, right-click the device and select 

Properties

 in the context menu.

•  On the 

General

 Tab, change the 

Thermocouple type

 in 

the drop down menu.

•  Apply

 these changes, there will be a prompt to reset the 

device, select 

yes

.

Alarm Settings

To change the settings for the alarm:
1.  Select 

Alarm Settings

 from the 

Device Menu

 in the 

MadgeTech Software. A window will appear allowing to 
set the high and low alarms and warning alarms.

2.  Press 

Change

 to edit the values.

3.  Check 

Enable Alarm Settings

 to enable the feature 

and check each high and low, warn and alarm box 
to activate it. The values can be entered in the field 
manually or by using the scroll bars.

4.  Click 

Apply

 to write these changes to the device. To 

clear an active alarm or warn, press the 

Clear Alarm

 or 

Clear Warn

 button.

5.  To set an alarm delay, enter the duration of time into 

the 

Alarm Delay

 box in which the readings can be 

outside of the alarm parameters.

Multiple Start/Stop Mode Activation

To start device:

 Press and hold the pushbutton for  

5 seconds, the green LED will flash during this time. The 
device has started logging.

To stop the device:

 Press and hold the pushbutton for  

5 seconds, the red LED will flash during this time. The 
device has stopped logging.