Product User Guide | 4
UltraShock
PRODUCT OVERVIEW
Humidity
The UltraShock utilizes polymer type humidity sensor and
a high-tech Gore® vent material at the entrance holes to
the internal temperature and humidity sensor chamber.
The vent equalizes pressure, reduces condensation and
also blocks entry of liquids and other contaminants.
If the unit has been exposed to very high levels of humidity
for extended time, it will return to normal measurement
accuracy by storing it in normal room humidity
(approximately 50% RH) conditions for several days.
Temperature
The UltraShock utilizes a silicon temperature sensor and
is calibrated for accuracy. When measuring humidity
and temperature, in addition to shock, please note that
there will be a delay in response time to humidity and
temperature measurements as the UltraShock processes
shock events before other parameters.
Pressure
The UltraShock pressure sensor is designed to measure
barometric pressure from very low altitudes to those
altitudes found in aircraft, up to high altitude balloons.
While specified to 300 mbar, it will continue to report data
to 100 mbar. This equates to approximately 53,000 feet.
The pressure sensor is contained within the same sealed
chamber as the temperature and humidity sensors and is
also protected by the Gore® vent.
SOFTWARE INSTALLATION
MadgeTech 4 Software
The MadgeTech 4 Software makes the process of downloading and reviewing data quick and easy, and is
free to download from the MadgeTech website.
1. Download the MadgeTech 4 Software on a Windows PC by going to
.
2. Locate and unzip the downloaded file (typically you can do this by right clicking on the file and selecting
Extract
).
3. Open the
MTInstaller.exe
file.
4. You will be prompted to select a language, then follow the instructions provided in the MadgeTech 4 Setup Wizard to
finish the MadgeTech 4 Software installation.