background image

 Product User Guide  |  9

UltraShock

SPECIFICATIONS

TEMPERATURE

Sensor

Semiconductor

Range

-20 ºC to +60 ºC

Resolution

0.1 ºC

Accuracy

±0.5 °C (+5 ° to +60 °C)

HUMIDITY

Sensor

Capacitive Polymer

Range

0 %RH to 95 %RH

Resolution

0.1 %RH

Calibrated 

Accuracy

±3 %RH (±2 %RH typical at 25 ºC)

Specified 

Accuracy Range

+20 °C to +40 °C; 25 %RH to 80 %RH

PRESSURE

Sensor

Semiconductor Strain Gage

Range

100 mbar to 1300 mbar

Resolution

0.05 mbar

Calibrated 

Accuracy

±1.5 mbar at 25 ºC; at 750 mbar

SHOCK

Accelerometer 

Type

MEMS Semiconductor

Acceleration  

Range (g)

±15 g

±100 g

±300 g

Acceleration 

Resolution (g)

0.02 g

0.05 g

0.2 g

Calibrated  

Accuracy (g)

±0.3 g

±2.0 g

±6.0 g

Sampling Rate

1000 Hz

Accelerometer 

Freq. Resp.

>1000 Hz (15 g)

>500 Hz (100 g, 300 g)

Reading Rate

1024 Hz to 5 minutes for shock, selectable 

in software. Temperature, pressure and 

humidity sampled approx. every  

2 seconds at intervals shorter than  

2 seconds. Otherwise, sampled at the 

reading rate.

GENERAL

Memory

3,593,358 readings

(598,893 per channel, all channels in use)

Start Modes

Software programmable immediate start  

or delay start, up to 6 months in advance

Real Time 

Recording

May be used with PC to monitor and 

record instantaneous measurements in 

real time

Password 

Protection

An optional password may be 

programmed into the device to restrict 

access to configuration options. Data may 

be read out with the password

Calibration

Digital calibration through software

Calibration Date

Automatically recorded within device

Battery Type

Internal Lithium Ion pack, charger 

included

Battery Life

20-90 days dependent on user settings 

(reading rate, trigger level, peak or 

instantaneous)

Data Format

Date and time stamped gravities (g and 

mg), temperature (°C, °F, K, °R), humidity 

(%RH, mg/ml water vapor concentration), 

pressure (PSIA, inHg, mmHg, bar, atm, 

Torr, Pa, kPa, MPa)

IP Rating

IP64

Time Accuracy

10 seconds/month (at 0 ºC to 50 ºC)

Computer 

Interface

USB-C cable required (included); 1MBaud

Software

Windows XP SP3 or later

Software 

Compatibility

Standard Software version 4.2.15.0 or later

Secure Software version 4.2.14.0 or later

Operating 

Environment

-20 ºC to +60 ºC,

0 %RH to 95 %RH non-condensing

Dimensions

3.4 in x 1.7 in x 1.3 in

(86 mm x 43 mm x 33 mm)

Weight

8 oz (227 g)

Enclosure

Anodized aluminum

Approvals

CE compliant

EMC Directive 2014/30/EU

RoHS Directive 2011/65/EU

Battery Warning:

 

FIRE, EXPLOSION AND SEVERE BURN HAZARD.  DO NOT SHORT CIRCUIT, CRUSH, PENETRATE, INCINERATE OR DISASSEMBLE.  AVOID 

TEMPERATURES ABOVE THE MAXIMUM OPERATING TEMPERATURE OF THE PRODUCT.  DISPOSE OF PROPERLY.  CHARGE ONLY WITH THE PROVIDED 

MADGETECH CHARGER, OR FROM OTHER USB POWER SOURCE VIA THE MADGETECH PROVIDED CABLE.

Specifications subject to change. See MadgeTech’s terms and conditions at 

madgetech.com

.

Summary of Contents for UltraShock

Page 1: ...ck Data Logger PRODUCT USER GUIDE TABLE OF CONTENTS 2 Product Overview 4 Software Installation Guide 5 Device Operation 8 Product Maintenance 9 Specifications 10 Dimensional Drawings To view the full...

Page 2: ...occur when the UltraShock is mounted by two bolts directly to a solid metal flat surfaced object of significant mass For various applications the rigid bolted down mounting may not be optimal or prac...

Page 3: ...SB connected and adequate battery charge level Flashes every 15 seconds maximum Indicates low battery charge level Logging Data with USB communication Flashes every 15 seconds Indicates normal operati...

Page 4: ...ts before other parameters Pressure The UltraShock pressure sensor is designed to measure barometric pressure from very low altitudes to those altitudes found in aircraft up to high altitude balloons...

Page 5: ...pending upon your start method 6 Disconnect the data logger from the interface cable and place it in the environment to measure Note The device will stop recording data when the end of memory is reach...

Page 6: ...s will only sample data at the logging rate For instance if a 30 second logging rate was selected then only a single data point is captured at each 30 second instant The potential to lose valuable dat...

Page 7: ...100 samples are acquired at the selected logging rate The trigger sample count should be chosen carefully because it may gather data points for which we have no interest For instance if 100 were enter...

Page 8: ...cell pack Please note the safe handling and storage recommendations in the specifications section Inspection The UltraShock may be inspected visually for any signs of damage especially to the gray gas...

Page 9: ...nto the device to restrict access to configuration options Data may be read out with the password Calibration Digital calibration through software Calibration Date Automatically recorded within device...

Page 10: ...21 8 mm 1 72 43 7 mm 0 20 5 1 mm 2X 0 20 5 1 mm THRU 4X R0 25 R6 4 mm APPLIQUE COVER GASKET BATTERY PACK RH AND PRESSURE SENSOR PORT TYPE C USB CONNECTOR 3 00 76 2 mm RED LED INDICATOR GREEN LED INDIC...

Reviews: