Choosing this setting will lead to a Warning – confirmation screen for making sure
that it was not accessed by mistake, going forward is possible with they Yes function key:
a)
b)
Drawing 42.
Electrochemical sensors calibration: a) warning screen, b) a list of all possible IR
sensors.
Similarly to IR sensors, all sensors available for calibration will be visible in the
sensor list and are marked with the badge. The calibration screen will be displayed
after the sensor is chosen and function key Enter is pressed.
Drawing 43.
Electrochemical sensor calibration screen.
Electrochemical sensor calibration process is the same as in case of IR sensors.
The calibration curve for the electrochemical sensor is in fact a straight line, that
can be set simply with a use of only two calibration points (for zero gas and reference
gas).