Drawing 38.
Info screen.
This screen contains basic information about the analyser like: software version,
serial number, number of the last saved report, and a list of all installed in the analyser
sensors.
5.3.6.6.2. Control list
This option is only available for the authorised madur services.
5.3.6.6.3. IR sensors calibration
Attention!
IR sensors calibration performed by untrained member of staff can
cause loss of correct analyser's setting and in consequence make
further measurement impossible.
Madur company strongly discourages changing settings in this
menu without a proper training!!!
IR sensor calibration from the analyser is only a small correction of 1 or 2
calibration points – the factory set shape of the curve will not change, this operation will
only move the curve on the graph.
37