15
Technical Specifications
Supply voltage and frequency:
230 V 50 Hz
Refrigerant fluid:
R-600a / 40 g
Nominal power:
125 W
Maximum current at compressor start-
up:
10 A
Nominal useable volume:
250 L
Maximum no. of doses:
450 blisters /400 tubes/ 450 jars
Temperature measurement resolution:
0.1ºC
Temperature range:
15ºC to 40ºC, +/-0.5ºC
External dimensions:
540 mm x 585 mm x 1440 mm
(height)
Unladen weight:
45 kg
Operating conditions:
For use within laboratories
Ambient temperature: 5-40ºC
Maximum altitude: 2000 m
Relative humidity: 80% at 31ºC
up to 50% at 40ºC
Maximum voltage fluctuations in
electricity supply: +/- 10%
Category II transient overvoltage