Aethalometer HAP 

 

 

 

___________________________________________________________________________________________________________ 

User's manual 

 Version 1.0                                                  March 2021                                                      6/17

 

 

1.2

 

Intended use and description of the equipment. 

 

The AE33/AE43 internal suction pump draws the sample air stream into the inlet connector 
and  through  the  analytical  filtration  area.    The  pump  speed  is  under  closed-loop  software 

control to stabilize the sample flow rate to the desired set point. Two flow sensors measure the 

airflow in the two sampling streams.  Their data is used in the BC calculation as well as for the 

closed-loop flow stabilization. 

The internal suction pump in combination with the flow sensors can supply a constant sample 

flow rate of 5 SLPM up to altitudes of about 3000 m above sea level.  When operated at higher 

altitudes, the sample flow rate provided by the internal pump will decrease. 

 

Figure 3. 

Reported flow as a function of altitude for Aethalometer internal suction pump (red line) and 

Aethalometer High Altitude Booster Pump (blue line)

 

 

The Aethalometer High Altitude Booster Pump will extend the rated performance of 

AE33/AE43 instruments to altitudes of 5000 m above sea level.  

 

4

4.5

5

5.5

6

6.5

7

7.5

8

0

1000

2000

3000

4000

5000

6000

Re

p

o

rted

 flow

 [

SLPM]

Attitude [m above sea level]

Aethalometer HAP

AE33 internal suction pump

Summary of Contents for Aerosol Aethalometer AETH-HA-PUMP

Page 1: ...______________________________________________________________________________________________ User s manual Version 1 0 March 2021 1 17 Aethalometer High Altitude Booster Pump User s Manual Version 1...

Page 2: ...contents of this manual with their own names trademarks and logos for branding purposes and for the distribution of the Magee Scientific Aethalometer High Altitude Booster Pump The text may not be mo...

Page 3: ...cking the equipment 5 1 2 Intended use and description of the equipment 6 1 3 Technical specification 7 2 IDENTIFICATION SAFETY NOTES and LABELS 8 2 1 Equipment identification label 8 2 2 Safety notes...

Page 4: ...provides a constant sample flow rate of 5 SLPM stabilized by the internal mass flow sensors from sea level to an altitude of approximately 3000 m At higher altitudes while the analytical optics of the...

Page 5: ...de Booster Pump AETH HA PUMP are as follows External AC DC power supply VES150PS12 Control cable HAP 1 99 004 Power supply cord B88F306 EU M6115 US M6117 China M5118 UK Black conductive sample line tu...

Page 6: ...culation as well as for the closed loop flow stabilization The internal suction pump in combination with the flow sensors can supply a constant sample flow rate of 5 SLPM up to altitudes of about 3000...

Page 7: ...Width 14 cm Height 19 cm Weight 4 90 kg Environmental Operating Conditions Indoor use only environmental protection IP X0 ordinary equipment Temperature range 10 40 C Relative humidity range non conde...

Page 8: ...VERY CAREFULLY BEFORE OPERATING THE EQUIPMENT 2 1 Equipment identification label 2 2 Safety notes and labels Equipment operation Read this User s manual BEFORE operating the equipment INCORRECT equip...

Page 9: ...lied with the equipment see below Fire and explosion NEVER install the equipment in explosion risk areas and never use the equipment near flammable substances Equipment overheating ALWAYS assure the e...

Page 10: ...n an even surface next to the Aethalometer The maximum distance between AETH HA PUMP and AE33 AE43 should not exceed 1 5 m The equipment is supplied with two meters of black conductive sample line tub...

Page 11: ...with the Booster Pump Connect them as shown in Figure 5 Figure 5 Connect AETH HA PUMP 12 VDC 150 W connector to power supply with External AC DC power supply adapter VES150PS12 Connect AETH HA PUMP C...

Page 12: ...mum lower flow of the HAP Figure 6 Flow range set valve is located on the rear panel NOTE If the valve is fully open the minimum lower flow of the HAP may exceed the maximum flow of the internal AE33...

Page 13: ...SLPM throughout the entire extended operating range Figure 8 Figure 8 By partly closing the flow range set valve the AE33 AE43 can provide a constant sample flow rate of 5 SLPM throughout the entire e...

Page 14: ...regularly We recommend the following schedule for regular maintenance Maintenance activity Maintenance period Cleaning with dry cloth As needed Inspect the sample line tubing M4995 Monthly Inspect Ext...

Page 15: ...ervice Departments through our web sites Service Support www mageescientific com or www aerosol eu Aerosol d o o Kamni ka 39a SI 1000 Ljubljana SLOVENIA EUROPE tel 386 1 4391 700 support aerosol eu ww...

Page 16: ...contents of this manual with their own names trademarks and logos for branding purposes and for the distribution of the Magee Scientific Aethalometer High Altitude Booster Pump The text may not be mo...

Page 17: ...Aethalometer HAP ___________________________________________________________________________________________________________ User s manual Version 1 0 March 2021 17 17...

Reviews: