background image

SAFETY NOTES

Please pay attention to additional safety notes as described in detail in the User Manual, Chapter 2, 

„SAFETY NOTES AND LABELS“

.

THE INSTRUMENT

8“ color 

Touch screen

Status indicator LED's

USB ports

Sample inlet

Air exhaust

Fan cover

CF card cover

USB, COM ports

Network connection

front view

back view

ON/OFF switch

(rear panel and behind front door)

INSTALLATION

Installation example:
Aethalometer AE33, Sample Stream Dryer, cyclone and MET station 

Sample Stream Dryer, Cyclone, MET station - OPTIONAL

Any water intake may damage  the instrument. 
Damage due to water intake 

is not covered by the warranty! 

1) Remove 

red

protective caps from the

instrument.
2) Attach the sample inlet line using
fittings and adapters. Avoid sharp bends.
3) Switch the instrument on (make sure 
to switch on TWO switches, one on rear panel, 
one behind front door).
4) 

Use the size-selective PM2.5 cyclone inlet 

to exclude 

dust and large particles. Clean the cyclone periodically!

5) Use the Sample Stream Dryer especially in cases of
high indoor ambient humidity.
6) Avoid direct discharge of air-conditioner over 
sample lines, or insulate them.  (This may cause 
water condensation in the tubing).
7) Install insect screen/water trap (provided
with the instrument) close to the rear of the instrument. 
The cup must be oriented downwards.

The instrument should preferably be installed indoors. If outdoors, it must be protected from the 
weather (sun, rain, wind). It should be operated at ambient temperature from 10 to 40

°

C. If it is 

installed in a small enclosure, this must be ventilated by a fan. Avoid placing the instrument or its 
sampling lines in the discharge from air-conditioners.

Mag

ee

 Scie

n

ti

fic 

Ae

thal

om

et

er 

AE33 

QUI

CK 

ST

AR

GUI

DE

SAMPLE STREAM DRYER

VACUUM PUMP

AE3

3

Avoid direct discharge of air-
conditioner over sample lines.

Reviews: