Electrical Connections
The inspection unit as supplied is internally wired for operation from a power source as designated on the
Magnaflux Data Plate (located on headstock end of inspection unit). When connecting the unit all local codes
should be followed. External wiring directions for operation on the designated voltage are as follows:
1. Open rear panel located on the headstock end, or Optional disconnect, to gain access to input power
terminal block.
WARNING:
Turn off and lock-out the external disconnect switch (not normally supplied)
before opening the rear panel. Failure to disconnect incoming power could
result in severe electrical shock or death.
2. Connect external power source to the terminal block on the rear of the inspection unit, or the top
connectors in the Optional disconnect. Refer to Table 1-1 for recommended dual-element time delay
fuse size and power source conductor size.
WARNING:
Magnetic particle baths can carry electrical current. Since operators will come
in contact with the bath during operation, proper grounding of the inspection
unit is essential. Failure to ground the inspection unit properly may result in an
electrical shock hazard. Check the ground connection with an ohmmeter. A
reading of 0 ohms denotes an acceptable ground connection.
In Figure 2-1 above, the rear headstock side panel has been removed. This is the location for the electrical
connection. Figure 2-2 above is a closer view of where the connection is located. The black wires on the right
side of the High Voltage label in Figure 2-2 would be the wires that the customer installs.
In Figure 2-3 above, the copper lug shown is the ground lug. The top punch out hole in the frame is for the
customer to run the wires through and connect to the terminal block located behind the High Voltage label.
Figure 2-2
Figure 2-1
Figure 2-3
Summary of Contents for D-2000 Series
Page 1: ...D 2000 Series Revised March 2020 Page 1 0 D 2000 SERIES ...
Page 9: ...D 2000 Series Revised March 2020 Page 1 8 PAGE LEFT BLANK INTENTIONALLY ...
Page 20: ...D 2000 Series Revised March 2020 Page 2 11 PAGE LEFT BLANK INTENTIONALLY ...
Page 33: ...D 2000 Series Revised March 2020 Page 3 13 PAGE LEFT BLANK INTENTIONALLY ...
Page 40: ...D 2000 Series Revised March 2020 Page 4 7 PAGE LEFT BLANK INTENTIONALLY ...
Page 45: ...D 2000 Series Revised March 2020 Page 5 5 PAGE LEFT BLANK INTENTIONALLY ...
Page 48: ...D 2000 Series Revised March 2020 Page 6 3 PAGE LEFT BLANK INTENTIONALLY ...