Table 2-1. Recommended Fuse Size
D-2060 and D-2060 MOD 2.5 AC
Input Volts AC
Phase
Full Load (Amps)
Line Fuse (Amps)
Wire Size (AWG)
Ground wire
208
230
380
415
460
575
3
3
3
3
3
3
160
160
100
90
80
70
125
125
80
70
60
60
#2
#2
#6
#6
#6
#6
Same as line
or larger
D-2100 and D-2100 MOD 4.0
Input Volts AC
Phase
Full Load (Amps)
Line Fuse (Amps)
Wire Size (AWG)
Ground Wire
208
230
380
415
460
575
3
3
3
3
3
3
400
400
245
220
200
160
300
300
200
175
150
125
#0
#0
#2
#2
#4
#4
Same as line
or larger
D-2060 MOD 6AC and D-2100 MOD 6AC
Input Volts AC
Phase
Full Load (Amps)
Line Fuse (Amps)
Wire Size (AWG)
Ground wire
208
230
380
415
460
575
N/A
3
3
3
3
3
3
N/A
575
350
325
300
200
N/A
350
200
200
150
100
N/A
2/0
0
0
2
2
Same as line
or larger
Note:
The equipment can be fused based on less than the maximum amperage draw due to the duty cycle.
Refer to NEC Code Section 630.31 Ampacity of Supply Conductors and Table 630.31(A) (2) Duty Cycle
Multiplication Factors for Resistance Welders (NPFA National Electric Code 2011). Some areas refer to 2
phase instead of 1 phase. The incoming power goes directly to an isolation transformer so the unit will operate
identically on 1 or 2 phase power.
Summary of Contents for D-2000 Series
Page 1: ...D 2000 Series Revised March 2020 Page 1 0 D 2000 SERIES ...
Page 9: ...D 2000 Series Revised March 2020 Page 1 8 PAGE LEFT BLANK INTENTIONALLY ...
Page 20: ...D 2000 Series Revised March 2020 Page 2 11 PAGE LEFT BLANK INTENTIONALLY ...
Page 33: ...D 2000 Series Revised March 2020 Page 3 13 PAGE LEFT BLANK INTENTIONALLY ...
Page 40: ...D 2000 Series Revised March 2020 Page 4 7 PAGE LEFT BLANK INTENTIONALLY ...
Page 45: ...D 2000 Series Revised March 2020 Page 5 5 PAGE LEFT BLANK INTENTIONALLY ...
Page 48: ...D 2000 Series Revised March 2020 Page 6 3 PAGE LEFT BLANK INTENTIONALLY ...