D-2000 Series
Revised March 2020
Page | 2-5
normal operation. Figure 2-8 shows the front cover removed however you have to remove the entire cover to
do any maintenance.
NOTE:
The following procedure should be performed by a qualified maintenance technician.
Ammeter Accuracy Verification
1. With the unit electrically installed, the compressed air supply connected, and the inspection hood
(optional) installed, connect the foot switch to the foot switch receptacle or the optional Dual Palm to
the Dual Palm receptacle on the left side of the unit. Refer to Figure 2-9 to locate both receptacles.
Figure 2-8
Figure 2-7
Foot Switch Receptacle
Dual Palm Receptacle
Figure 2-9