D-2000 Series
Revised March 2020
Page | 4-5
5. Press PUMP button to turn on the pump. The bath circulating pump motor should start. There should
be no leakage of bath solution. Operate the spray nozzle(s) and check for bath discharge. Release the
nozzle lever and check that the bath discharge stops. Press PUMP button to turn off the pump.
6. Connect the foot switch or Optional Dual Palm to the corresponding receptacle on the headstock end
of the machine.
7. If an Optional hood has been installed, connect the line plug to the HOOD outlet on the headstock end
of the machine.
8. Connect the black light to the convenience outlet on either end of the machine. The black light should
come to full brilliance in about 5 minutes. To turn off the black light, the plug must be removed from the
convenience outlet.
9. Press LIGHT button to turn on the hood white light. Press again to turn light off.
CIRCULAR (CONTACT) AND LONGITUDINAL (COIL) MAGNETIZATION CHECK
1. Move the coil against the headstock so it will not interfere. Lock in place.
2. Unlock and move tailstock to a position where the digital test meter kit shunt, noted in Table 4-5, can
be supported by the tailstock and headstock shelves. Lock tailstock.
3. If equipped with Auto-bath, press the AUTO BATH button off. The button is dimly lit when off and
bright when on.
4. Position the digital test meter kit shunt between the headstock and tailstock. The terminal with the red
test lead should be closest to the headstock. Lock the tailstock in position. Connect the shunt cable to
the test meter shunt input as directed in the test kit instructions.
5. Press the foot switch or Optional Dual Palm to firmly clamp the shunt in position between the
headstock and tailstock.
6. Press CONTACT button.
7. Set the amperage using the CONTACT current control or use the AMPERAGE SETPOINT on the HMI
to a desired value.
8. Hold the digital test meter as far from the machine as the leads will permit. Press and release the MAG
push-button (see test meter instructions). The green current assurance indicator should light during
the “shot,” and the current should be indicated on both the Digital Ammeter and the digital test meter.
Each ammeter reading should correspond within ±10% or 50 amps of the machine rating. {MOD AC
(Optional) repeat for AC}.
Summary of Contents for D-2000 Series
Page 1: ...D 2000 Series Revised March 2020 Page 1 0 D 2000 SERIES ...
Page 9: ...D 2000 Series Revised March 2020 Page 1 8 PAGE LEFT BLANK INTENTIONALLY ...
Page 20: ...D 2000 Series Revised March 2020 Page 2 11 PAGE LEFT BLANK INTENTIONALLY ...
Page 33: ...D 2000 Series Revised March 2020 Page 3 13 PAGE LEFT BLANK INTENTIONALLY ...
Page 40: ...D 2000 Series Revised March 2020 Page 4 7 PAGE LEFT BLANK INTENTIONALLY ...
Page 45: ...D 2000 Series Revised March 2020 Page 5 5 PAGE LEFT BLANK INTENTIONALLY ...
Page 48: ...D 2000 Series Revised March 2020 Page 6 3 PAGE LEFT BLANK INTENTIONALLY ...