Airway adapters are single use. The re-use may result in cross contamination.
10.18.6.14 ICO
2
alarm
To test the iCO
2
alarm, connect a capnograph to the ventilator. Assembly the respiratory circuit to use
with the capnograph and blow the airway adapter so that the ventilator monitors the capnograph parameters.
To test the high iCO
2
alarm, set the alarm’s limit iCO
2
below the monitored in the equipment.
•
Airway adapters are single use. The re-use may result in cross contamination.
10.18.6.15 Heart rate alarm
To test the heart rate alarm, connect an oximeter to the ventilator. Assembly the respiratory circuit and
connect the oximeter in the finger to monitor the oximeter parameters. To test the high heart rate alarm, set
the alarm’s upper limit heart rate below the monitored in the equipment. To test the low heart rate
alarm, set
the alarm’s lower limit heart rate above the heart rate monitored in the equipment.
10.18.6.16 SpO
2
alarm
To test the SpO
2
alarm, connect an oximeter to the ventilator. Assembly the respiratory circuit and
connect the oximeter in the finger to monitor the oximeter parameters. To test the SpO
2
alarm, set the alarm’s
limit SpO
2
below the monitored in the equipment.
10.18.7 Battery test
To check the battery status, enter the Battery menu, where the voltage and battery status information
will be displayed. To test the operation of the battery, disconnect the equipment from the electrical power and
check that the remaining battery power is displayed in percentage. Wait a few minutes, reconnect the machine
to the electrical power and check that the battery is recharged.
10.19 Performance Specifications
Table 40 - Performance specifications