145
NOTE 4 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Fields in the vicinity of RF wireless communication equipment
The Fleximag Max enclosure interface was tested as specified in the table below using the test methods
specified in IEC 61000-4-3
Band
Freq. test
Modulation
Test level
[ MHz ]
[ MHz ]
[ V/m ]
380 to 390
385
Pulse, 18 Hz
27
430 to 470
450
FM, 1 kHz,
Deviation from ± 5kHz
28
704 to 787
710
745
780
Pulse, 217 Hz
9
800 to 960
810
870
930
Pulse, 18 Hz
28
1.700 to 1.990
1.720
1.845
1.970
Pulse, 217 Hz
28
2.400 to 2.570
2.450
Pulse, 217 Hz
28
5.100 to 5.800
5.240
5.500
5.785
Pulse, 217 Hz
9
•
The EMISSION characteristics of this equipment make it suitable for use in industrial
areas and hospitals (ABNT NBR IEC / CISPR 11 class A). If used in a residential
environment (for which ABNT NBR IEC / CISPR 11 class B is normally required), this
equipment may not provide adequate protection for radio frequency communication
services. The user may need to take mitigation measures, such as relocating or
reorienting the equipment.