43
Item
Procedure
6
Verify that the breathing circuit is firmly connected and is suitable for the
patient.
7
Verify the tight connection of oxygen and compressed air hoses.
8
Verify that the inlet pressure is within the specified range.
•
Input pressures exceeding the specified limit can damage the
equipment.
•
For inlet pressures less than 250 kPa, the maximum flow shall
be 120 L/min.
9
Check the power cord connection, if applicable.
The ventilator can be used on battery for up to 210 minutes continuously,
under normal ventilation conditions.
•
If LOW BATTERY alarm occurs, IMMEDIATELY connect the
ventilator to the electrical network. If this is not possible,
provide another suitable means of ventilatory support and
then DISCONNECT the equipment from the patient.
10
If all items are OK, then the equipment is ready for use.
•
Perform all verification procedures before each use.
•
If any problem is found, correct it BEFORE USING THE DEVICE.
•
If the problem cannot be resolved immediately, contact qualified technical assistance.