6.2 Restricted menu
This menu is password-only accessible and allows you to make advanced settings, calibrations and
autotests, and allows you to view system information, battery and status.
6.2.1 Restricted menu
6.2.1.1 Patient
Clicking this button will return you to the patient screen.
6.2.1.2 Autotest
In this menu, you can perform the tests, and cancel and view the results individually. This menu
displays the air pressure, O
2
information, date and time of the last test performed.
6.2.1.3 Calibration
In this menu, it is possible to perform the calibration of the distal flow sensor, expiratory valve and
oxygen cell. This menu displays the air pressure, O
2
information, date and time of the last test performed.
To perform the calibrations, occlude the breathing circuit, select the items you want to calibrate, and
press the calibrate button.
6.2.1.1
6.2.1.8
6.2.1.2 6.2.1.3
6.2.1.6 6.2.1.7
6.2.1.4 6.2.1.5