6.3.8 Expiratory hold button
Pressing this button pauses the Expiratory time for the time set in the ADVANCED menu. This function
allows for extended expiration time maneuvers (prolong expiration time). During the execution of this maneuver,
the apnea alarm is not displayed. Only the message "Exp Pause" is displayed.
6.3.9 O
2
+
button (O
2
flush)
When this button is pressed, an O
2
flush is performed with the concentration and time set in the
ADVANCED menu. During this time, the high FiO
2
alarm will be inhibited. This feature can be used for pre-
and post-aspiration of airway secretion procedures and is available in all ventilatory modes. Use with an
aspiration catheter may be performed in any mode and does not require a specific adjustment.
6.3.10 Nebulizer / TGI button
By pressing this button, nebulization or TGI is performed according to the setting in the RESOURCES
menu.
6.3.11 Graphic freeze button
When you press this button, the graphics are frozen.
6.3.12 Menu access area
Pressing this button displays the following menus: Layout, Resources, Maneuver, Trend, Calibration,
Advanced, Status and Engineering mode.
6.3.12.1 Layout
Pressing this button displays the available graphic layouts: 2 waveforms, big numbers, monitor, 3
waveforms, lung and 2 loops.
2 waveforms
Big numbers
A
B
C
C
B