82
connected to the data communication port.
•
Failure to implement the communication protocol will result in failure to send data to
other equipment.
10.4.4.1 Protocol used for data communication with external devices
The Ethernet port may be used to share ventilator data, such as adjusted parameters, monitored
parameters, waveforms, and alarm logging for electronic health recorders. The data has on average a delay
of 8 seconds between the instant of data generation and the data output connector.
To send data to other equipment, the IT network must be scalable, with high availability and low data
propagation delay.
The required network settings include a Dynamic Host Configuration Protocol (DHCP) enabled
network server, so that the SEMP receives a valid Internet Protocol (IP). Communication is done through the
TCP protocol in the IT network. For communication with the electronic health recorder, an appropriate
communication protocol shall be implemented. To obtain the implementation guide for the communication
protocol, contact MagnaService.
The information transitions from the following form: Fleximag Max sends the data to the responding
client it received. The customer can ask questions or an operational request for Fleximag Max that immediately
responds or confirms the request.
•
To obtain the implementation guide for the communication protocol, contact
MagnaService.
•
This implementation shall be performed in a network with the characteristics described
in 10.4.4.1 by an IT specialist.
•
IT network failures to provide the required features may lead to delays in data
communication or incorrect, incomplete or corrupted data transmission, resulting in
incorrect information for the user.